Microsoft Word - 299-2475-1-LE-rev


ACTA IMEKO 
ISSN: 2221‐870X 
November 2016, Volume 5, Number 3, 76‐80 

 

ACTA IMEKO | www.imeko.org  November 2016 | Volume 5 | Number 3 | 76 

Surface analytical model and sorption artifact designing 
method 

Xiao‐ping Ren
1,2
, Jian Wang

1
, Ombati Wilson

3
,Yong Wang

4
,Bai‐fan Chen

4,5
, Chang‐qing Cai

1
 

1
Mass Laboratory, Division of Mechanics and Acoustics, National Institute of Metrology, Beijing, P.R.China 

2
Department of R&D Management, National Institute of Metrology, Beijing, P.R.China 

3
Kenya Bureau of Standards, Nairobi, Republic of Kenya 

4
College of Information Science & Engineering, Central South University, Changsha, P.R.China 

5
Department of Computer Science and Engineering, Texas A&M University, USA 

 

 

Section: RESEARCH PAPER  

Keywords: mass measurement; surface sorption corrected; mass dissemination; optimization algorithm 

Citation: Xiao‐ping Ren, Jian Wang, Ombati Wilson ,Yong Wang, Bai‐fan Chen, Chang‐qing Cai, Surface analytical model and sorption artifact designing 
method, Acta IMEKO, vol. 5, no. 3, article 12, November 2016, identifier: IMEKO‐ACTA‐05 (2016)‐03‐12 

Section Editor: Marco Tarabini, Politecnico di Milano, Italy 

Received December 15, 2015; In final form August 1, 2016; Published November 2016 

Copyright: © 2016 IMEKO. This is an open‐access article distributed under the terms of the Creative Commons Attribution 3.0 License, which permits 
unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited 

Funding:  This  work  was  supported  by  National  Natural  Science  Funds  of  China  (51405459),  National  Science  and  Technology  Support  Program 
(2011BAK15B06) and Special‐Funded Program on National Key Scientific Instruments and Equipment Development (2012YQ090208). 

Corresponding author: X.P. Ren, e‐mail: renxp@nim.ac.cn 

 

1. INTRODUCTION 

The unit of mass, the kilogram, is the last of the seven base 
units of the International System of Units (SI) to be defined 
according to an invariant of nature rather than a material 
artefact [1]. Both the watt balance method and Avogadro 
method are making the new definition under vacuum 
conditions whereas the current definition of the unit, from the 
International Prototype Kilogram (IPK), is realized and 
maintained in air.  

After the new definition, it is still necessary to consider 
traceability from the new realization in vacuum to the current 
working standards which are always maintained in air [2]. This 
is to make an indirect link between air and vacuum mass 
measurements by measuring mass in vacuum and then 
characterizing the absorption layers of contaminants during the 
process  of  transferring  from  vacuum  to  air  [3].  During this  

 
 
 

transition, the mass of the standard is significantly affected by a 
sorption phenomenon. This phenomenon is caused by 
atmospheric gases and humidity, which subsequently lead to 
loss of stability of the mass value of the standard(s) [4]. In 1973, 
Takayoshi studied the problem of surface water on metal 
artefacts [5]. R. Schwartz wrote a series of papers on adsorption 
isotherms in air [6] and sorption phenomena in vacuum [7]. 
Additional studies focused on Pt/Ir [8], stainless steel [6], [7], 
silicon [9] and also other materials (Au [8], [10], [11]).  

For mass dissemination, there are many parameters to be 
considered in designing the mass standard, like height, diameter, 
volume, surface area and mass value of the weight. The 
characteristics of a measurement instrument are also important 
such as the size of the mass comparator, electronic weighing 
capacity and volume measurement instrument. In this paper, an 
adaptive algorithm which can be used to optimize the design of 

ABSTRACT 
Mass standards with alternative shapes are difficult to design due to the number of complex parameters. An analytical model based 
on  surface  sorption  experiments  is  presented  to  study  adsorption.  This  model  is  based  on  an  optimization  algorithm  that  is 
conceptualized to help to design the best sorption artefacts. Experimental artefacts, cylinder‐weight and stack‐weight, were of the 
same volume but different surface areas. This algorithm in essence determines the optimum surface of the artefact. After machining 
the artefact, surface sorption measurements were carried out. A sorption experiment was done by transferring the artefact from air to 
a vacuum. Then the surface sorption model was set up which represented the relationship between sorption coefficient η, time t and 
relative humidity h. Logarithmic models were used to fit the variation of sorption coefficient η per relative humidity h with time t.



 

ACTA IMEKO | www.imeko.org  November 2016 | Volume 5 | Number 3 | 77 

a mass standard based on a surface analytic model of mass 
standards is described. 

2. RECENT RESEARCH ON SURFACE ARTIFACTS 

One kilogram weights with different surface areas and same 
nominal mass values are normally required. Samples of different 
materials (stainless steel, silicon material and platinum-iridium) 
have to be selected. 

Figure 1 shows an example of designing a stack of weights 
made of platinum-iridium assembled using a spacer (height of 1 
cm and diameter of 2 mm). The volumes, masses and surface 
areas of the spacers were added to the overall values for the 
weight stack. For some applications of this kind of stack 
weight, when taking volume measurement, one disc should be 
placed into the liquid and then the spacer is placed on its 
surface. Then another disc is put onto the spacer until all the 
separate discs and spacers are immersed in the volume 
measurement liquid. During this process, however, the stack of 
weights may fall down from the magazine and lead to process 
and system failure. 

Another kind of stack weight was composed of a number of 
discs [6]. These discs were tightened on carrying rods (D: 10 
mm). The discs and carrying rods were polished by Häfner 
Company. For this kind of stack weight, there was a hole in the 
middle of each disc where the rod was screwed in. These discs 
cannot be totally seamlessly fixed together (as shown in Figure 
2). Thus whenever this stack weight is submerged into the 
liquid for volume measurement, the volume value would be 
affected by this gap. A similar design method was described by 
Beer [13].  

Another kind of artefact, shown in Figure 3, consisted of 
twelve discs separated with pieces of wire and held together 
with a thin rod. This particular artefact was gold coated with a 6 
μm thick layer with an aim of determining the characterizations 
of the gold surfaces. 

In Section 3 we present a mathematical model that, when 
optimized through computational methods, provides for the 
best design parameters of the artefacts to be machined. 

3. INTEGRAL SORPTION ARTEFACT 

In mass measurements, the main uncertainties are due to air 
buoyancy correction and surface sorption correction. In order 
to improve the accuracy of the surface analytical model of a 
mass standard, the most important thing is to reduce the 
influence of the buoyancy correction as much as possible.  

Two prototype models of 1 kg stainless sorption artefacts 
are shown in Figure 4 and Figure 5. The model represented by 
Figure 4 is the classical prototype, and it is in cylindrical form 
with height and diameter being equal. The other model, shown 
in Figure 5, is in the form of a stack and the discs are separated 
by a rod. This rod is not however separated from the disc; it is a 
monolith of stainless steel. 

The surface area (S) and volume (V) of a cylinder weight 
(Figure 4) are shown in (1) and (2), where r, D and H represent 
the radius, diameter and height of weight respectively. 

 
cylinder

2
22 2 2 2

2 2
            

   
D D

S r rH H , (1)
 

cylinder

2

2
    
 

D
V H .

 
(2) 

The dimensions of the stack prototype (shown in Figure 5) 
are as follows: outer circle’s radius and height being R2 and H2 
respectively; inner circle’s radius and height being R1 and H1 
respectively. The total surface area and volume of 4-level stack 
prototype are shown in (3) and (4): 

2 2
stack 1 1 2 2 2 16 8 8 6S R H R H R R       , (3)  

 2 2
stack 1 1 2 23 4V R H R H    . (4) 

Generally, the 4-level stack prototype above is an example of 
designing the sorption artefact. Different levels can be adopted 
during design of the sorption artefact depending on the 
sorption effect. The normal formulae for calculating the volume 

 
Figure 3. 1 kg gold‐plated copper buoyancy artefact. 

Figure  1.  Designing  of  a  platinum‐iridium  weight  set  by  Johnson  Matthey 
and NPL [12]. 

 
Figure 2. The gap between disc and rod. 

        
Figure 4. Cylinder prototype.           Figure 5. Discs of stack prototype. 



 

ACTA IMEKO | www.imeko.org  November 2016 | Volume 5 | Number 3 | 78 

and surface area of stack weight with different levels n are 
shown in (5) and (6): 

2 2
1 1 2 2( 1)V n R H n R H    ,                                                         (5) 

2 2
1 1 2 2 2 12( 1) 2 2 2( 1)S n R H n R H n R n R         .  (6) 

In this research, the two models of mass standards are 
expected to be of the same volume but different surface areas 
i.e. (2) and (4) should be equal. Thus, the air buoyancy 
correction applied during the comparison process will be 
minimized. The large difference in surface areas between (1) 
and (3) is better for the mass measurement during transferring 
from vacuum to ambient. 

4. PARAMETER  SEARCHING  METHOD  BASED  ON  THE 
OPTIMIZATION ALGORITHM 

For the same material with ideal density ρ, and nominal mass 
value (1 kg), both height H and radius r of cylinder prototype 
can be calculated by (2). For sorption purposes, let (5) equals 
the cylinder volume. Thus, there are four parameters which 
determine the volume of the n-level stack, i.e. R1, R2, H1 and 
H2. With surface area of cylinder scylinder being also known, these 
four parameters can therefore be varied until the maximum 
difference between scylinder and sstack is obtained. However, this 
design scheme cannot neglect the usage of both mass and 
volume measurement equipment. All final artefacts therefore 
can be put into these instruments and their mass and volume 
can be measured accordingly.  

During determining the volume and surface area of the 
cylindrical prototype, a 9-level stack was considered. This 
design scheme had the following limitations: 

(a) R2 < 50 mm (size of the weighing magazine); 
(b) height of weight less than 90 mm (volume 

measurement instrument space); 
(c) volume of 9-level stack equal to the volume of 

cylindrical prototype i.e. Vstack = Vcylinder ; 
(d) surface area of 9-level stack > Surface area of 

cylindrical weight i.e. Sstack > Scylinder;  
(e) mass difference between cylindrical and 9-level stack 

prototype less than or equal to 1.5 g (i.e. electronic 
weighing capacity). If the mass difference exceeds this 
limitation, the equipment malfunctions or gives 
misleading results; 

(f) machining precision equal to 0.01 mm, so the precision 
of 2 decimal points for each parameter is enough. For 
example, if the optimal height of the weight H is 
53.3725 mm, the value is rounded to 53.37 mm.  

In this study, there were six parameters [D, H, R1, R2, H1 and 
H2] and 6 constraint conditions being considered in the 
sorption artefact. There are several approaches for solving this 
kind of multi-object optimization. Matlab has an optimization 
toolbox with the function “fgoalattain”, whose graphical user 
interface is shown in Figure 6. The following parameters were 
determined and programmed: start point, goals, weights, linear 
inequalities, bounds for variables. Objective function and 
nonlinear constraint function were written into a .m file, which 
described the formula listed from (1) to (6). More details are 
given in Appendix I and II. 

Start point and final results are respectively shown in Tables 
1 and 2. When the optimization algorithm is initially running 
from the start point, it executes 13 iterations and the algorithm 
makes the judgement whether volume and surface satisfy the 

requirement; otherwise, the algorithm is re-executed from the 
current result until the maximum difference of surface areas is 
obtained and volume difference is close to zero. Optimization 
trends of surface area, volume and iteration numbers of 
optimization algorithm at different start points are shown in 
Figure 7. Figure 7(b) shows that the optimization sequence of 

Table 1. Start Point for the optimization algorithm (unit: mm).

H R R1 R2 H1 H2 

0.01 0.01 0.01 0.01 0.01 0.01 

 
 
Table 2. Optimization result for the start point (unit: mm). 

H R R1 R2 H1 H2 

53.27 26.64 4.04 50 9.50 1.62 

Figure 7. The trend of surface area, volume and the  iteration numbers of 
optimization algorithm at different start points. 

Figure 6. Interface for Global Optimization Toolbox in Matlab 2011.  

10 20 30 40 50
0

5

10

15

x 10
4

Optimization Sequence
                 (a)                  

S
u
rf

a
ce

 A
re

a
(m

m
2
)

 

 

0 20 40
108

110

112

114

116

118

120

122

Optimization Sequence
                     (b)                  

V
o

lu
m

e
(c

m
3
)

 

 

0 10 20 30 40 50
0

20

40

60

Optimization Sequence
                 (c)                  

A
lg

o
ri

th
m

 I
te

ra
ti
o
n
 N

u
m

b
e

r

 

 

Cylinder

9 Level Stack

9 Level Stack

Cylinder



 

ACTA IMEKO | www.imeko.org  November 2016 | Volume 5 | Number 3 | 79 

volume may have unsatisfactory results (peaks). The curve (red 
line) in Figure 7(c) represents the variation trends, and shows 
that the algorithm reached stability after 52 optimization 
sequences. From this particular optimization results, the 
maximum surface area difference and volume difference were 
determined to be 133699.71 mm2 and 0 cm3 respectively. 

5. SURFACE SORPTION EXPERIMENT 

After machining the artefacts (in 2014), their surfaces were 
thoroughly cleaned (using alcohol) to remove any contaminate 
e.g. oil and dust. They were subsequently set to stabilize for one 
year within laboratory ambient condition. These artefacts were 
also cleaned before the sorption measurements in air and 
vacuum.  

5.1. Surface measurement in air 

During this experiment, the masses of the 9-level stack 
weight and cylindrical weight described in Section 4 were 
measured using the Mettler-Toledo M-one system. 
Measurements were performed under normal air condition; 
their masses were measured from Sept. 10th to Sept. 14th 2015 
and the mass differences thereof computed. This measurement 
procedure was again repeated at the same condition a month 
later for two days (i.e. Oct. 15th and Oct. 16th). Results of 
sorption measurements in air are as shown in Figure 8, in which, 
ΔI, Δm, h, t and ρ respectively represent balance indication, 
mass difference, relative humidity, temperature in the 
measurement chamber and air density. The differences between 
the two series of tests were analysed. 

5.2. Surface measurement in vacuum 

Artefacts were transferred from air to vacuum (TM and PM 
sensor installed in mass measurement system: M-one); the 
pressure ranged between 7×10-5 Pa and 2×10-3 Pa. The 
sorption was measured from Oct. 20th to Oct. 23th, 2015; results 
are shown in Figure 9. Figure 9(a) shows that the measurement 
values were stable at the end of the tests; the last mass 
difference was 1.2207 mg. Combined with the measurement 
results in air condition, variation of the relative mass per area A 
could be determined by mass comparison of the sorption 
artefacts.  

5.3. Relation between sorption coefficient and humidity 

 In 1994, Schwartz [6] determined the sorption coefficient η as 
η=Δm/ΔA+η0. To minimize the effect of η0, the current study 
ensured that the artefacts were thoroughly cleaned before 
performing measurements, and hence η0 was assumed to be 
zero. From the scatter plot of Figure 10, the horizontal axis and 
vertical axis respectively represent real time t (in hours), and 
sorption coefficient per relative humidity η/h (unit: 
μg/cm2/RH %). Sorption coefficient and relative humidity 
were fit with the aid of the logarithmic model using the 
following the function in (7):   

 f(η)= (4.3×10-4+ 3.5×10-5ln (0.43 t +1)) ·h .  (7) 

The results obtained from (7) were similar to the results of 
Schwartz’s model of the sorption coefficient. However, unlike 
Schwartz’s model, the present model introduced humidity into 
the function and also obeyed the logarithmic rule. 

This study was congruent with earlier research findings on 
negligible influence of the effect of roughness condition and 
temperature of weight surface [6], and did not, therefore, 
investigate their effects on the sorption behaviour of weight. 

6. CONCLUSIONS 

In order to provide a practical approach on disseminating 
the redefined kilogram, realized in vacuum to the mass scale at 
ambient, processes such as air to vacuum transferring for 

Figure 8. Result of the sorption measurement in air stage  
(a) Variation  of  the  ΔI  reading  from  the  balance  with  measuring  times

(broken line: long term gap).  
(b) Variation of mass difference Δm with measuring times.  
(c) Variation of relative humidity h with measuring times.  
(d) Variation  of  temperature  t  in  the  measurement  chamber  with

measuring times.  
(e) Variation of air density ρ with measuring times.  Figure 9. Measuring the mass difference under vacuum condition. 

0 20 40 60 80 100 120 140

-0.84

-0.835

-0.83

∆
I(

in
te

rv
a
l)

measurement in air

0 20 40 60 80 100 120 140

-0.8

-0.795

-0.79

∆
m

(m
g
)

0 20 40 60 80 100 120 140

35

40

45

h
(R

H
%

)

0 20 40 60 80 100 120 140

20.5

20.6

20.7

t(
�

)

0 10 20 30 40 50 60 70 80 90 100 110 120 130 140

1.19

1.191

1.192

Measuring times

ρ
(m

g
/c

m
3
)

(a)

(b)

(d)

(e)

(c)

Sep.10th~14th
Oct.15th~16th

0 10 20 30 40 50 60 70 80 90
-1.222

-1.22

-1.218

-1.216

-1.214

-1.212

(
)

∆
I

in
te

rv
a
l

measurement in vacuum

0 10 20 30 40 50 60 70 80 90
0

2

4

6

8
x 10

-3

T
M

2
(h

P
a
)

0 10 20 30 40 50 60 70 80 90
0

0.5

1

1.5
x 10

-3

measuring times

P
M

(h
P

a
)

Oct 20th

(a)

(b)

(c)

Oct 23th

t 
(°

C
) 



 

ACTA IMEKO | www.imeko.org  November 2016 | Volume 5 | Number 3 | 80 

standards must be studied. This transfer process and the results 
thereof are often affected by adsorption and desorption. In this 
paper, a new surface analytical model combining artefact 
standards designing method is presented. The model is based 
on an optimization algorithm which considers important 
parameters such as suitable design of weights (i.e. cylindrical 
weight and stacks of discs of weight). These parameters could 
help the metrologists in improving the accuracy of mass 
measurements. 

The preparation of the sorption artefact and the sorption 
measurement has also been described. Sorption coefficient drift 
was be described by a logarithmic function that includes the 
effects of the humidity and time. 

In this experiment, the detailed value of roughness was not 
given, but the relationship between time, humidity and sorption 
coefficient was presented. Further research may be conducted 
in future on artefacts to determine their drift for a long time in 
air condition. 

 

APPENDIX I 

function f = weight_design(x) 
density = 8421;    
n = 9;      
Pi=3.14159265;    
S_cylinder = 2*Pi*x(2)^2+2*Pi*x(2)*x(1);     
V_cylinder = Pi*x(1)*x(2)^2/1000;    
S_stack=2*(n-1)*Pi*x(3)*x(5)+2*n*Pi*x(4)*x(6)+2*n*Pi*x(4) 
^2 -2*(n-1)*Pi*x(3)^2; 
V_stack = ((n-1)*Pi*x(3)^2*x(5) + n*Pi*x(4)^2*x(6))/1000;                
f(1) = -S_stack + S_cylinder;     

APPENDIX II 

function [c,ceq] = nonlconstr(x) 

E1 = 0.5;  
density = 8421;  
n = 9;  
Pi=3.14159265;  
c(1) = -x(1)*x(2)^2 +(1000*(1000000-E1))/Pi/density; 
c(2) = x(1)*x(2)^2-(1000*(1000000+E1))/Pi/density; 
c(3)=(1-n)*x(5)*x(3)^2-n*x(6)*x(4)^2-(1000*(E1-1000000))/ 

Pi/density; 
c(4)=(n-1)*x(5)*x(3)^2+n*x(6)*x(4)^2-(1000*(E1+1000000))/ 

Pi/density; 
ceq = Pi*x(1)*x(2)^2-(n-1)*Pi*x(5)*x(3)^2-n*Pi*x(6)*x(4)^2; 

REFERENCES 

[1] S. Davidson, J. Berry, Z. Silvestri, et al, “Addressing the 
requirements for the practical implementation and ongoing 
maintenance of the redefined kilogram,” Proc. Of IMEKO 22nd 
TC3, 12th TC5 and 3rd TC22 International Confer-ences, Feb.3-
5, 2014, Cape Town, Republic of South Africa. 

[2] Mise en pratique of the definition of the kilogram. 
http://www.bipm.org/en/si/new_si/mise-en-pratique.html, 
2013. 

[3] P. J. Abbott, R.C. Dove. “Progress on a vacuum-to-air mass 
calibration using magnetic suspension to disseminate the plank-
constant realized kilogram,” Proc. Of IMEKO 22nd TC3, 12th 
TC5 and 3rd TC22 International Conferences, Feb.3-5, 2014, 
Cape Town, Republic of South Africa. 

[4] P Fuchs, K Marti, S Russi. “New instrument for the study of ‘the 
kg, mise en pratique’: first results on the correlation between the 
change in mass and surface chemical state,” Metrologia, 2012, 
V49, pp. 607–614. 

[5] Takayoshi, Shuiti, et al, “Coulometric Micro-determination of 
surface water on various metals and glasses and of hydrogen in 
beryllium metal,” Materials Transactions, V 14, 1973, pp. 396-
400. 

[6] R. Schwartz. “Precision Determination of Adsomtion Lavers on 
Stainless Steel Mass Standards by Mass comparison and 
Ellipsometry. Part I: Adsorption Isotherms in Air,” Metrologia, 
V 31, 1994, pp. 117-128. 

[7] R. Schwartz. “Precision Determination of Adsorption Layers on 
Stainless Steel Mass Standards by Mass Comparison and 
Ellipsometry Part II Sorption Phenomena in Vacuum,” 
Metrologia, V31, 1994, pp.129-136. 

[8] P. Fuchs, K. Marti , S. Russi. “Materials for mass standards: long-
term stability of Pt/Ir and Au after hydrogen and oxygen low-
pressure plasma cleaning,” Metrologia, V 49, 2012, pp. 615–627. 

[9] M. Borys, M. Mecke, U. Kuetgens, et al. “The growth of the 
oxide layer on silicon spheres and its influence on their mass 
stability,” Proc. Of IMEKO 22nd TC3, 12th TC5 and 3rd TC22 
International Conferences, Feb.3-5, 2014, Cape Town, Republic 
of South Africa. 

[10] P. Fuchs. “Low-pressure plasma cleaning of Au and PtIr noble 
metal surfaces,” Applied Surface Science, V 256, 2009, pp.1382–
1390. 

[11] K. Marti, P. Fuchs , S. Russi. “Cleaning of mass standards: II. A 
comparison of new techniques applied to actual and potential 
new materials for mass standards,” Metrologia, V 50, 2013, pp. 
83–92. 

[12] S. Davidson, S. Brown, J. Berry.”A report on the potential 
reduction in uncertainty from traceable comparisons of 
platinum-iridium and stainless steel kilogram mass standards in 
vacuum”, NPL Report CMAM 88, 2004. 

[13] W. Beer, W. Fasel, E. Moll, P. Richard, et al. “The METAS 1 kg 
vacuum mass comparator – adsorption layer measurements on 
gold-coated copper buoyancy artefacts,” Metrologia, V 39,2002, 
pp. 263-268. 

 

Figure 10. Relation between time t and η/h. 

0 10 20 30 40 50 60 70

4.3

4.4

4.5

4.6

4.7

4.8

x 10
-4

time t(hour)

s
o

rp
tio

n
 c

o
e

ffi
c
ie

n
t 
p

e
r 

re
la

tiv
e

 h
u

m
id

ity
(μ

g
/c

m
2
/%

)

 

 

η /h vs t

logarithmic model