Study on alignment test based on ASTM E1012 and Chinese guideline


ACTA IMEKO | www.imeko.org December 2020 | Volume 9 | Number 5 | 159 

ACTA IMEKO 
ISSN: 2221-870X 
December 2020, Volume 9, Number 5, 159 - 162 

 
 

STUDY ON ALIGNMENT TEST BASED ON ASTM E1012 AND CHINESE 

GUIDELINE 
 

Xiaobing Zhao1, Wansheng Li2, Minghui Fei3, Xiaolan Zheng4, William Huang5 

 
1 Wuxi Institute of Metrology and Testing, Wuxi, China, zxb@wxmtc.com  

2 Shandong Institute of Metrology, Jinan, China, liwansheng@sdim.cn  
3 Wuxi Institute of Metrology and Testing, Wuxi, China, fmh@wxmtc.com  

4 Yunnan Institute of Measuring and Testing Technology, Kunming, China, zhengxlan06@163.com  
5 GTM China Office, Shanghai, China, william.huang@gtmchina.cn  

 

 

Abstract 

This paper describes the study about alignment 

test on testing machine. It is based on the standards 

of ASTM E1012-14 (Standard Practice for 

Verification of Testing Frame and Specimen 

Alignment Under Tensile and Compressive Axial 

Force Application) and Chinese calibration 

guideline. The strain-gauge specimens of ASTM 

and reference extensometer of Chinese guideline 

were both tested on one testing machine. The test 

results were analyzed and compared. The 

traceability about both methods were also discussed. 

Meanwhile the amplifier was calibrated before test.    
It is the first time in China to performance this 

comparison test. 

Keywords: alignment test, ASTM E1012, 

Chinese alignment guideline 

1. INTRODUCTION 

The alignment on testing machine is very 

important to obtain the right test results of material. 

The standard of ASTM E1012-14 [1] and Chinese 

guideline [2] [3] both descript how to test the 

alignment. But the methods are different, and the 

philosophy of traceability are also not same. To 

prove both methods reliable on testing the 

alignment at one testing machine, it is necessary to 

do a comparison test to see if there will be big 

tolerance by these two standards. Therefore, two 

laboratories in China, both own both equipment of 

ASTM E1012 and Chinese alignment guideline, 

decide to do the comparison test in 2019.  

ISO 23788:2012 (Metallic materials – Verifi-

cation of the alignment of fatigue testing machines) 

[4] also describes about alignment test, but it is very 

similar as ASTM E1012. The most difference 

between ISO 23788 and ASTM E1012 is: there are 

total 12 pieces of strain-gauges on three levers at the 

transducer according to ASTM, but there are only 8 

strain-gauges on two levers according to ISO. 

The details and difference between ASTM 

E1012 and Chinese guideline are: the specimen, the 

procedure, the calculation, the classification 

(deviation) and the traceability. They are described 

in section 2 to section 5. 

2. THE EQUIPMENTS  

According to ASTM E1012, there are four 

strain-gauges at same lever but different position on 

alignment specimen (so-call alignment transducer). 

Total 12 or 8 pieces of strain-gauges on 3 or 2 levers 

at the transducer. As Figure 1: 

 
Figure 1: Strain-gauge transducer 

The equipment to be performance alignment test 

consist: an alignment specimen with 12 pieces of 

strain-gauges, amplifier of 12 channels, the software. 

They are called alignment tool kit by the product 

from ZwickRoell, a Germany testing machine 

manufacturer. As Figure 2. 

The alignment specimen details (sharp, 

dimension) as Figure 3. 

http://www.imeko.org/
mailto:zxb@wxmtc.com
mailto:liwansheng@sdim.cn
mailto:fmh@wxmtc.com
mailto:zhengxlan06@163.com
mailto:william.huang@gtmchina.cn


ACTA IMEKO | www.imeko.org December 2020 | Volume 9 | Number 5 | 160 

 
Figure 2: Alignment tool kit from ZwickRoell 

 
Figure 3: Alignment specimen 

The amplifier is type quantumX MX1615B. 
Excitation voltage is 10V and full-bright are set in 

the software. 12 channels of this amplifier are used 

to be connected to the alignment specimen. 

According to Chinese guideline JJG 475-2008: 

verification of electric testing machine, and JJG 

556-2011:verifications of uniaxial fatigue testing 

machine [2] [3], the alignment kit is twins’ 

extensometers and specimens, as Figure 4. 

 
Figure 4: twins’ extensometers on specimen 

The twins’ extensometers mean there are two 

extensometers: one is attached on the left side of 

specimen and another is on the right side. The 

extensometer is used for measuring the elongation 

of the specimen during tensile test. Theoretically the 

elongation of both sides of specimen should be same 

if the alignment of the testing machine is perfect. 

The specimen’s specification is descripted as in 

Table 1 and Figure 5. 

Fmax of testing 
machine in kN 

diam.d 
in mm 

gauge 

length Lo 

in mm 

L in 
mm 

material 

[100---600) 10 100 130 
steel 

No.45 

(30---100) 10 100 130 
alloy 

steel 

Table 1: the dimension of the specimen 

 

Figure 5: the drawing of the specimen 

3. PROCEDURE  

We use a ZwickRoell testing machine Z050 to 

do test. The capacity of the machine force is 50KN. 

The classification of this testing machine is class 0.5. 

 

3.1 According to ASTM E1012-14: 

1. clamp the alignment transducer at the top grip, 

zero the signal, clamp the lower grip. 

Apply a small force to see all signal ok and then 

remove the force. 

2. increase the force and apply it to 4 %, 6 % and 

8 % of the capacity of the testing machine, that 

means 2 kN, 3 kN, and 4 kN. Record the 

measurement value. And calculate all the data 

automatic by software. 

(The original is to apply the force to 10%, 20% 

and 40% of the machine capacity, here we apply 4%, 

6% and 8% of the machine capacity) 

3. remove the force to zero. Change the position 

of the alignment transducer from position 0° to 180° 

and repeat the step 2. 

4. remove the force to zero. Change the position 

from 180° back to 0°, and repeat step 2. 

 

3.2 According to Chinese guideline: 

1. clamp the specimen on both grips of testing 

machine 

2. apply the force of 0.5 % of the machine and 

attach the twin’s extensometers at the position of 1 

and 3 on the specimen as Figure 6. 

Apply the force to 1 % of the machine and zero 

the signal of both extensometers. Apply to the force 

to 10 % of the machine. That means the pre-apply. 

Do it three times. 

3. remove the force to zero. Apply the force to 

1 % of the machine and zero the signal of both 

extensometers. Increase the force to 4 %, 6 % and 

8 % of the machine. Record all the data. 

(The original is to increase the force to 4% ,  6%, 

8%  and 10% of  the machine capacity, here we 

apply 4%, 6% and 8% of the machine capacity) 

http://www.imeko.org/


ACTA IMEKO | www.imeko.org December 2020 | Volume 9 | Number 5 | 161 

 

4. change the extensometers to positions 2 and 

position 4. Same procedure as step 2 and 3. 

 

 

Figure 6: the positions for twins’ extensometers attached 

4. DATA ANALYSISI 

According to ASTM E1012, the Cylindrical 

strain-gauged alignment specimen use following 

equations:(As Figure 7) 

 
Figure 7: the equations of PB 

PB is the percent bending, which is the value of 

the alignment. 

From the test 3.1, all data were collected. Here 

we show the data from the middle lever of the strain-

gauge, as in Table 2. 

In ASTM E1012, the original “record data points 

at 10%, 20% and 40% of the force transducer range 

or testing machine capacity”, but here we only apply 

the force to 4%, 6% and 8%, and the best value of 

PB is 10.7 at 4kN (8% of the machine capacity of 

50K), because we try to make the same force apply 

as Chinese calibration guideline. 

 

 

 

Force Position PB (%) 
PB 

(Ave) 

2 kN middle 

0° 25.3 

21.5 180° 16.89 

0° 22.3 

3 kN middle 

0° 17.42 

14.3 180° 11.62 

0° 13.94 

4 kN middle 

0° 14.08 

10.7 180° 8.98 

0° 9.12 

Table 2: the data of the middle lever of the strain-

gauge 

 

According to the Chinese guideline, the value of 

e is the alignment specification of the testing 

machine. It uses following equations: (As Figure 8) 

 
Figure 8: the equations of e 

 

From the test 3.2, we collect the data of e as in 

Table 3. 

 

Force e (%) 

2 kN 10.3 

3 kN 7.3 

4 kN 5.4 

Table 3: the data by test 3.2 
 

Test results of test 3.1 and test 3.2 are compared 

as in Table 4. 

 

Force PB (%) e (%) 

4 % of 

machine 
2 kN 21.5 10.3 

6 % of 

machine 
3 kN 14.3 7.3 

8 % of 

machine 
4 kN 10.7 5.4 

Table 4: the results compared 

http://www.imeko.org/


ACTA IMEKO | www.imeko.org December 2020 | Volume 9 | Number 5 | 162 

The PB results are almost exactly double the e 

results. But it should be no coincidence because the 

classification is different as discussion later. 

5. DISCUSSION 

The classification of alignment by ASTM E1012: 

and Chinese calibration guideline is different, as in 

Table 5. 

ASTM E1012 

Classification 

PB 

(%) 
 

Classification 

of testing 

machine 

e 

(%) 

5 5 
 

0.5 12 

8 8  1 15 

10 10  2 20 

Table 5: the different classification. 
 

The standard of ASTM only talks about the 

alignment classification, instead of the machine. 

The Chinese guideline takes care the machine must 

fulfil the requirement of the alignment. 

The data from test 3.1 does not apply enough 

force on the strain-gauge specimen according to 

ASTM E1012. Therefore, it cannot give decision 

that PB of 10.7 does not fulfil class 10. 

The data from 3.2 test shows that the machine 

full fill class 0.5, because the alignment of e is 

10.3%, which is within the tolerance of 12%. 

Additional topic need taken care is the 

traceability. In ASTM E1012, there mentioned: 

“8.1.1 Calibration of strain-gaged alignment 

transducers is not required by this standard. 

Traceable national standards do not generally exist 

for such calibrations. However, great care should be 

taken in the manufacture of strain-gage alignment 

transducers used for the determination of alignment” 

That means only the amplifier can be calibrated. 

We did calibration as in Figure 9 for this. The 

standard signal is created by a device. 

The deviation between standard signal and 

amplifier (mV/V) is less than -0.3%. Its fine. The 

details data as shown in Table 6. 

standard 
amplifier 

mV/V 

amplifier 

μm/m 
deviation % 

0 0.0003 0.4  - 

0.25 0.2499 254.6 -0.04  

0.5 0.4997 509 -0.06  

0.75 0.7485 763 -0.20  

1 0.9977 1017 -0.23  

Table 6: calibration data of amplifier. 

 
Figure 9: The standard signal (above), and the amplifier 

to be calibrated (below) 
 

Concerning the calibration on twins’ 

extensometers of Chinese, its easy: calibrate the two 

extensometers one by one. Each is done by a master 

extensometer. They can be traceability to the 

primary standard of length. That mean all the 

Chinese alignment kit can be calibrated.  

6. SUMMARY 

The main difference between the standard of 

ASTM E1012-14 and Chinese calibration guideline 

about alignment are classification, tool kit hardware, 

and the force apply. 

ASTM is more accuracy on alignment 

specificity of testing machine. But there is no 

solution for the whole traceability. 

The method of Chinese guideline is easier to 

handle and can be traceability. 

 Both methods are reliable to determine the 

alignment of the material testing machine. 

7. REFERENCES 

[1] ASTM E1012-14. Standard Practice for 
Verification of Testing Frame and Specimen 

Alignment Under Tensile and Compressive Axial 

Force Application. 

[2] JJG 475-2008: verification of electric testing 
machine, 

[3] JJG 556-2011. verifications of uniaxial fatigue 
testing machine.  

[4] ISO 23788:2012 (Metallic materials – Verification 
of the alignment of fatigue testing machines) 

 

 

http://www.imeko.org/