Oral Sciences n3


Original Article Braz J Oral Sci.
July/September 2009 - Volume 8, Number 3

Assessment of flexural strength of two
self-curing acrylic resins containing pigment

Vicente Telles 1, Raphael M. F. Brito1, Estela M. Jurach 2, Lincoln I. Nojima 3

1Undergraduate student, Department of Orthodontics, Federal University of Rio de Janeiro, Brazil
2 DDS, MS, Graduate student, Department of Orthodontics, Federal University of Rio de Janeiro, Brazil

3 DDS, MS, Associate Professor, Department of Orthodontics, Federal University of Rio de Janeiro, Brazil

Received for publication: June 29, 2009
Accepted: November 11, 2009

Correspondence to:

Lincoln I. Nojima
Universidade do Brasil - UFRJ -

Faculdade de Odontologia
Programa de Pós-graduação em Odontologia

Av. Professor Rodolpho Paulo Rocco, 325
Ilha do Fundão - Rio de Janeiro - RJ

Brasil - CEP: 21941-617
Phone: 2590-2727  / Fax: 21-2590-9771

E-mail: linojima@gmail.com

Abstract
Aim: To assess the flexural strength of orthodontic acrylic resins from two different manufacturers ( VIPI,

OrtoCril versus Clássico, OrtoClas) by comparing pigmented resins to colorless ones. Methods: Resins of

blue, yellow and green colors were studied. A total of 120 specimens were made and then divided into

groups of 15 elements each, all having the same dimensions. Next, they were kept in aqueous medium

until being subjected to mechanical testing. The flexural strength was tested in a universal test machine

(EMIC DL 10000) in which the specimens were subjected to a gradual load until fracture occurred. Results:

Pigmented resins had flexural strength values compatible with clinical use, being similar to those from

colorless ones, except for OrtoClas green-colored and yellow-colored resins, which showed greater flexural

strength. The OrtoClas green-colored resin was the most resistant to fracture (482.2 N), whereas the OrtoCril

colorless resin was the least resistant (368.4 N). All OrtoClas resins showed higher strength values compared

to OrtoCril resins of same color, except for the OrtoCril’s blue-colored resin, which presented higher flexural

strength than that of the other trademark. Conclusions: The use of pigments seems to have no effect

on decreasing the flexural strength of self-curing acrylic resins. Therefore, pigmented resins are compatible

with clinical use.

Keywords: acrylic resins, flexural strength, orthodontic appliances.

Introduction
Self-curing acrylic resins have been widely used in orthodontics for making plates for small

tooth movements and space maintenance, palatal disjunction appliances, retention plates,

and fixed inclined planes1. Either orthodontists or even general practitioners can make some of

these more simple appliances in order to prevent progression of malocclusions, which can

potentially require a more complex and prolonged treatment in the future2. Therefore, despite

being inefficient in some orthodontic treatments, these appliances still play an important role

in correcting malocclusions during deciduous and mixed dentition within each stage of the

craniofacial development3-4. Because of their limitations, these orthodontic appliances are

used only in the treatment of children and teenagers as the practitioner can rely on both bone

growth and eruptive tooth movement.

Esthetics is so valued today that the demand for orthodontic baseplates that are colored,

decorated or inscribed with designs has increased. Because these appliances can be fixed or

removable, it is extremely important that the patient is satisfied and accepts them so that he

or she can co-operate with the orthodontic treatment. In order to fulfill such a demand,

pigmented acrylic resins has been increasingly accepted in the dental market as their preparation

is the same as that regarding colorless resins and there is no need to add pigment during the

laboratorial phase to obtain a colored resin. However, no study assessing the resistance of

these materials after pigmentation is available in the literature.

Braz J Oral Sci. 8(3):137-140



Although acrylic resin is widely used, this material can easily

fracture due to its low resistance to impact, low flexural strength or

low fatigue strength5-6. Flexural strength is a physical property that

determines the material’s resistance to bending. When a flexural force

is applied, the material suffers an elastic deformation followed by

plastic deformation and eventually fractures 7. Another great

disadvantage of acrylic resins is the rapid loss of esthetic, physical

and mechanical properties within the oral medium because this

material absorbs and releases water8.

Orthodontic appliances tend to fracture due to both occlusive

forces and presence of metallic wires which promote dental movement

because it suffers some deformation during placement and removal,

thus leading to fatigue over the insertion areas of these wires9-10.

Because the pigments added to resins can bring impurities that

might react with free radicals, the polymerization reaction could be

compromised and consequently the physical properties of the material

changed. Therefore, the aim of the present study was to assess the

flexural strength of pigmented resins compared to colorless resins.

Material and methods
A total of 120 acrylic resin specimens (20.5 x 5.5 x 4.0 mm) distributed

into 8 groups (n = 15) were fabricated using in a powder/liquid mixing

ratio of 3:1, according to recommendations of the two manufacturers,

namely, OrtoCril® (VIPI, Pirassununga, SP, Brazil) and OrtoClas® (Artigos

Odontológicos Clássico Ltda., São Paulo, SP, Brazil). The specimens

were made within a silicone condenzation mold (Perfil, Vigodent, Rio

de Janeiro, RJ, Brazil) with internal dimensions of 21 x 6.2 x 4.0 mm

that served as a negative control. Polymerization occurred inside a

VH Softline pressure container (Midas Dental Products Ltda.) at

constant pressure of 17 pounds/in2 or 87.93 cm/Hg for 20 min. The

excess material was removed from the specimens by using

progressively 150, 400 600-grit sandpapers until obtaining the desired

dimensions, which were measured with a Starrett caliper (Figure 1).

After polishing the specimens were stored in containers with

water for 24 h so that residual monomers could be released. After this

period, water was replaced and the specimens remained immersed

until being subjected to mechanical testing.

OrtoClas specimens were indicated by the abbreviation Cl,

whereas OrtoCril specimens were indicated by Cr. These groups were

also identified in terms of color as follows: colorless (L), yellow (Y),

blue (B) and green (G).

The test consisted of gradually applying a force to each sepcimen

by using a universal test machine (EMIC DL 1000; São José dos

Pinhais, PR, Brazil) at a crosshead speed of 5 mm/min until fracture

occurred. The machine has three shafts in which the two inferior ones

serves to hold the sample and the superior one serve to apply force to

the centre of the sample. The three shafts have the same ray of 2.5

mm in order to avoid differences in the results. The center of the

specimen was determined by using a millimeter rule and the resulting

central point was marked with an OHP marker pen. The machine

recorded the force that resulted in fracture (Figure 2).

Statistical analysis was performed with the SPSS software for

Windows (v. 13.0) and using two-way ANOVA and Tukey’s test in

order to determine the mean flexural strength of each group and to

compare them regarding the trademarks. The control group consisted

of colorless self-curing acrylic resins because of the absence of pigments.

Means and standard deviations were obtained for a significant level

of 5% (Table 1).

Results
In the control group containing OrtoCril specimens, the fracture force

was not significantly higher than that of other specimens of same

trademark (Table 1 and Figure 3). With regard to OrtoClas resins,

containing yellow-colored and green-colored resin specimens had a

significantly higher flexural strength than the blue-colored resin and

colorless resin specimens (Table 1 and Figure 3).

Figure 2 – Flexural strength test schematic‘s representation.

Figure 3 – Box Plot comparing trademarks and colours of acrylic resins. Mean, SD, max, min.

Assessment of flexural strength of two self-curing acrylic resins containing pigment138

Braz J Oral Sci. 8(3):137-140

Figure 1 – Samples after polishing.



n Mean S D M i n i m u m M a x i m u m Statistics

OrtoCril Colorless 15 368.38 40.73 291.59 423.98 A

OrtoCril Yellow 15 401.37 32.03 348.01 448.56 A

OrtoCril Blue 15 406.17 46.29 334.60 494.92 A

OrtoCril Green 15 369.50 24.14 336.28 416.16 A

OrtoClas Colorless 15 397.42 52.04 312.82 465.31 A

OrtoClas Yellow 15 461.77 34.80 396.05 509.44 B

OrtoClas Blue 15 402.49 39.97 332.37 452.47 A

OrtoClas Green 15 482.17 20.81 448.00 512.79 B

Table 1 – Descriptive statistical analysis of the mean flexural strength values (N).

Different letters mean statistically significant difference (p < 0.05).

By comparing the trademarks, OrtoClas yellow-colored and

green-colored resins were significantly more resistant than the OrtoCril

resins of same colors. With regard to blue-colored and colorless resins,

no statistically significant differences were found between the

trademarks (Table 1 and Figure 3).

Discussion
Acrylic resins are composed of polymeric chains of polymethylacrylate

and monomers of methylmethacrylate with a small amount of ligant

agents1,10. Polymerization occurs by mixing polymer (powder) with

monomer (liquid). A macromolecule is formed through a series of

chemical reactions11-12. Any impurity existing in the monomer may

inhibit or delay polymerization if free radicals are involved, thus

impeding the stage called propagation. Chemically polymerized resins

have 3-5% free monomers. A greater amount of these residual

monomers implies a decrease in resistance and hardness of the

material1,11-13.

According to Rocha Filho et al.12, there is an amount of residual

monomer following resin polymerization that is released mainly during

the first 24 h. Orthodontic resins have higher levels of residual

monomers compared to the conventional ones. In order to avoid any

influence from the residual monomer released, which might interfere

with the final physical properties of the resin, the aqueous medium in

which the specimens were stored was replaced following that period

of time. Residual monomers, if in contact with oral mucosa, can

cause local or systemic tissue reactions due to their cytotoxic activity14-

15. Pressure during polymerization raises the amount of residual

monomer as well as compromise the powder-liquid mixing ratio,

mainly in orthodontics as the salt-and-pepper technique is widely

used to saturate the polymer with monomer1,15-16.

According to Rantala et al.10 and Keyf and Etikan8, acrylic resins

undergo a water-absorbing process depending on the medium in which

they are. The greater the amount of residual monomer, the greater is

the water absorption. Water absorption decreases the mechanical

properties of the material in the oral cavity, since water acts as a

plastifying agent penetrating into the spaces between the polymeric

chains and decreasing the secondary chemical bonds such as the Van

der Waals forces1,14. With the aim of meeting the reality, we decided to

store the specimens in aqueous medium before submitting them to

mechanical test.

It has been reported that the addition of acrylic fibers or acrylic

stain to the polymerized resin through microwaves did not affect the

transverse strength, and that both methods were found to be

esthetically and mechanically acceptable for clinical use17. The present

study also shows that addition of pigments to acrylic resins does not

reduce their physical properties.

Although the flexural strength values are compatible with

masticatory forces, Price18 emphasizes that the majority of fractures

in acrylic resin are not the result of fatigue because such fractures

generally occur outside the oral cavity due to accidental impacts (e.g.:

falls). Despite the fact that these accidental impacts are not classified

as a functional stress, many manufactures have considered the impact

strength as being relevant factor to classify a given resin as being

“virtually unbreakable” under impact18. In order to increase the fracture

strength of acrylic resin, several types of fibers have been used such as

aramid fibers, glass fibers, nylon fibers, and carbon fibers, all showing

favorable results9,19-20. Nevertheless, because such techniques are not

frequently used for making orthodontic baseplates, they were not

assessed in the present study.

The use of a pressure container is indicated because the resin

cannot come in contact with oxygen during polymerization. Therefore,

the reaction speed and degree of polymerization are higher than those

in polymerization under air atmosphere because oxygen reacts with

free radicals, which results in less porosities12-13,20. The presence of

porosity not only reduces the mechanical properties of the resin, but

also interferes with the cleaning of orthodontic baseplates by allowing

adhesion of substances and deposition of calculus, which may promote

inflammatory processes in the surrounding mucosa11,16,21. Due to this

factor, finishing and polishing procedures using sandpapers as well as

the use of a pressure device were shown to be extremely important.

However, further studies are needed, as the pressure was used on an

arbitrary basis.

In addition, it was raised the question on why the OrtoClas self-

curing acrylic resins of yellow and green colors had higher flexural

strength. Although no further investigation was carried out in the

present study, it is possible that the polymer might be accounted

because the specimens were made using the same monomer. Therefore,

the reasons why OrtoClas acrylic resins of yellow and green colors

have higher flexural strength compared to colorless resins could not

be fully understood. The differences in the flexural strengths obtained

in various studies may be explained by the different methods of

polymerization and storage of the specimens before test.

It may be concluded that the use of pigments seems to have no

significant effect on decreasing the flexural strength of self-curing acrylic

resins. Although two pigmented resins had higher flexural strength,

other colors had mean values compatible with clinical use and can be

used according to the patient’s needs. In addition, the use of pigmented

resins does not interfere with the construction of the orthodontic

appliance.

Assessment of flexural strength of two self-curing acrylic resins containing pigment 139

Braz J Oral Sci. 8(3):137-140



Acknowledgments
The authors acknowledge the financial support given by CNPq,

CAPES and FAPERJ

The authors acknowledge the financial support given by CNPq,

CAPES and FAPERJ

References
1. Faltermeier A, Rosentritt M, Müssig D. Acrylic removable appliances:

Comparative evaluation of different postpolymerization methods. Am J Orthod
Dentofacial Orthop. 2002; 131: 301.e16-22.

2. Melsen B. Removable orthodontic appliances. Dent Clin North Am. 1981; 25:
157-76.

3. Pauw G, Derweduwen K, Dermaut L. Are removable appliances obsolete?
Ned Tijdschr Tandheelkd. 2000; 107:151-4.

4. Ward S, Read MJ. The contemporary use of removable orthodontic appliances.
Dent Update. 2004; 31: 215-8.

5. Kanie T, Fujii K, Arikawa H, Inoue K. Flexural properties and impact strength of
denture base polymer reinforced with woven glass fibers. Dental Mater. 2000;
16: 150-8.

6. Oliveira AG, Panzeri H. Flexural and fatigue strength of chemically activated
resin in association with a hybrid fibre. Biosci J. 2004; 20: 103-12

7. Elias CN, Lopes HP. Materiais dentários – ensaios mecânicos. São Paulo: Santos;
2007.

8. Keif F, Etikan I. Evaluation of gloss changes of two denture acrylic resin materials
in four different beverages. Dental Mater. 2004; 20: 244-51.

9. Mullarky RH. Aramid fiber reinforcement of acrylic appliances. J Clin Orthod.
1985; 19: 655-8.

10. Rantala LI, Lastumäki TM, Peltomäki T, Vallittu, PK. Fatigue resistance of
removable orthodontic appliance reinforced with glass fibre weave. J Oral
Rehabil. 2003; 30: 501-6.

11. Mantzikos T, Epstein M. Interior surface sealant for acrylic appliances. J Clin
Orthod. 1998; 32: 152-3.

12. Rocha Filho R, Paula LV, Costa VC, Seraidarian PI. Evaluation of residual
monomer in autopolymerizing acrylic resins: spectroscopy analysis Rev Dent
Press Ortod Ortop Facial. 2007; 12: 96-104.

13. Anusavice KJ. Materiais dentários. 10. ed. Rio de Janeiro: Guanabara Koogan; 1998.
14. Urban VM, Machado AL, Oliveira RV, Vergani CE, Pavarina AC, Cass QB. Residual

monomer of reline acrylic resins. Effect of water-bath and microwave post-
polymerization treatments. Dental Mater. 2007; 23: 363-8.

15. Gonçalves TS, Schmitt VM, Thomas M, Souza MAL, Menezes LM. Citotoxicity of
two autopolymerized acrylic resins used in orthodontics. Angle Orthod. 2008;
78: 926-30.

16. Gonçalves TS, Spohr AM, Souza RM, Menezes LM. Surface Roughness of auto
polymerized acrylic resin according to different manipulation and polishing
methods. Angle Orthod. 2008; 78: 931-4.

17. Silva FAP, Silva TBP, Rached RN, Del Bel Cury AA. Effect of intrinsic pigmentation
on the flexural strength of a microwave-cured acrylic resin. Braz Dent J. 2002;
13: 205-7.

18. Price CA. The effect of cross-linking agents on the impact resistance of a linear
poly(methyl methacrylate) denture-base polymer. J Dent Res. 1986; 65: 987-92.

19. John J, Gangadhar SA, Shah I. Flexural strength of heat-polymerized polymethyl
methacrylate denture resin reinforced with glass, aramid, or nylon fibers. J
Prosthet Dent. 2001; 86: 424-7.

20. Chung K, Lin T, Wang F. Flexural strength of a provisional resin material with
fiber addition. J Oral Rehabil. 1998; 25: 214-7.

21. Compagnoni MA, Barbosa DB, Souza RF, Pero AC. The effect of polymerization
cycle on the surface roughness in microwave-processed denture base resin.
Rev Odontol UNESP. 2005; 34: 101-6.

Assessment of flexural strength of two self-curing acrylic resins containing pigment140

Braz J Oral Sci. 8(3):137-140