All articles are permanently available online to the public without restrictions or subscription fees. All articles are free to be used, cited, and distributed, on condition that appropriate acknowledgment is included. Authors are the copyright holders of their original contributions and grant the Canadian Online Publication Group (COPG) a license to publish the article and identify itself as the original publisher. CPOJ articles are licensed under the Creative Commons Attribution 4.0 International License. CPOJ Website: https://jps.library.utoronto.ca/index.php/cpoj/index Editorial Office: cpoj@online-publication.com ISSN: 2561-987X CPOJ is a member of, and subscribes to the principles of, the Committee on Publication Ethics (COPE). VOLUME 3, ISSUE 2 2020 RESEARCH ARTICLE Seth M, Hou W, Gyrates L.R, Galassi J.P, Limber E.M. Effect of transtibial prosthesis mass on gait asymmetries. Canadian Prosthetics & Orthotics Journal. 2020;Volume 3, Issue 2, No.5. https://doi.org/10.33137/cpoj.v3i2.34609 https://online-publication.com/wp/ http://creativecommons.org/licenses/by/4.0/ https://jps.library.utoronto.ca/index.php/cpoj/index mailto:cpoj@online-publication.com https://publicationethics.org/about/our-organisation https://publicationethics.org/about/our-organisation https://publicationethics.org/members/canadian-prosthetics-orthotics-journal https://doi.org/10.33137/cpoj.v3i2.34609 1 Seth M, Hou W, Goyarts L.R, Galassi J.P, Lamberg E.M. Effect of transtibial prosthesis mass on gait asymmetries. Canadian Prosthetics & Orthotics Journal. 2020;Volume 3, Issue 2, No.5. https://doi.org/10.33137/cpoj.v3i2.34609. RESEARCH ARTICLE EFFECT OF TRANSTIBIAL PROSTHESIS MASS ON GAIT ASYMMETRIES Seth M1*, Hou W2, Goyarts L.R1 , Galassi J.P1, Lamberg E.M1 1 School of Health Technology and Management, Stony Brook University, Stony Brook, NY, USA. 2 School of Medicine, Stony Brook University, Stony Brook, NY, USA. INTRODUCTION Individuals living with a transtibial amputation (TTA) typically present with an asymmetrical gait pattern characterized by a prolonged stance phase (temporal asymmetry)1-5 and greater loading on the intact side as compared to the prosthetic side (limb loading asymmetry).5-8 A prolonged asymmetrical gait may predispose the individual to poor health outcomes, such as, knee or hip osteoarthritis of the intact side,8,9 back pain,10 and an increased risk of falling.11 Further, walking with an asymmetrical gait pattern may attract unwanted attention towards the individual.12 While, inherent factors associated with a TTA, such as weak prosthetic side push-off force 13 and issues pertaining to load bearing of the residual limb,14 may contribute to gait asymmetry, certain prosthetic factors, such as mass, may also contribute to gait asymmetry. Currently, the impact of prosthesis mass on gait asymmetry is not well defined. Further exploration may be vital to identifying any potential negative consequences that OPEN ACCESS Volume 3, Issue 2, Article No.5. 2020 Journal Homepage: https://jps.library.utoronto.ca/index.php/cpoj/index ABSTRACT BACKGROUND: Individuals with transtibial amputation (TTA) typically walk with an asymmetrical gait pattern, which may predispose them to secondary complications and increase risk of fall. Gait asymmetry may be influenced by prosthesis mass. OBJECTIVE: To explore the effects of prosthesis mass on temporal and limb loading asymmetry in people with TTA following seven days of acclimation and community use. METHODS: Eight individuals with transtibial amputation participated. A counterbalanced repeated measures study, involving three sessions (each one week apart) was conducted, during which three load conditions were examined: no load, light load and heavy load. The light load and heavy load conditions were achieved by adding 30% and 50% of the mass difference between legs, at a proximal location on the prosthesis. Kinematic and ground reaction force data was captured while walking one week after the added mass. Symmetry indices between the prosthetic and intact side were computed for temporal (Stance and Swing time) and limb loading measures (vertical ground reaction force Peak and Impulse). FINDINGS: Following seven days of acclimation, no significant differences were observed between the three mass conditions (no load, light load and heavy load) for temporal (Stance time: p=0.61; Swing time: p=0.13) and limb loading asymmetry (vertical ground reaction force Peak: p=0.95; vertical ground reaction force Impulse: p=0.55). CONCLUSION: Prosthesis mass increase at a proximal location did not increase temporal and limb loading asymmetry during walking in individuals with TTA. Hence, mass increase subsequent to replacing proximally located prosthesis components may not increase gait asymmetry, thereby allowing more flexibility to the clinician for component selection. ARTICLE INFO Received: July 16, 2020 Accepted: October 12, 2020 Published: October 19, 2020 CITATION Seth M, Hou W, Goyarts L.R, Galassi J.P, Lamberg E.M. Effect of transtibial prosthesis mass on gait asymmetries. Canadian Prosthetics & Orthotics Journal. 2020;Volume 3, Issue 2, No.5. https://doi.org/10.33137/cpoj.v3i 2.34609 KEYWORDS Prosthesis, Mass, Gait, Lower limb amputation, Kinematics, Kinetics, Ground reaction forces, Transtibial amputation, Gait asymmetry. * CORRESPONDING AUTHOR: Mayank Seth, PhD Delaware Limb Loss Studies Lab, University of Delaware, Newark, USA. E-mail: mseth@udel.edu ORCID: https://orcid.org/0000-0003-3526-7058 https://doi.org/10.33137/cpoj.v3i2.34609 https://jps.library.utoronto.ca/index.php/cpoj/index https://doi.org/10.33137/cpoj.v3i2.34609 https://doi.org/10.33137/cpoj.v3i2.34609 mailto:mseth@udel.edu https://orcid.org/0000-0003-2933-4365 https://orcid.org/0000-0003-2933-4365 2 Seth M, Hou W, Goyarts L.R, Galassi J.P, Lamberg E.M. Effect of transtibial prosthesis mass on gait asymmetries. Canadian Prosthetics & Orthotics Journal. 2020;Volume 3, Issue 2, No.5. https://doi.org/10.33137/cpoj.v3i2.34609. ISSN: 2561-987X EFFECT OF TRANSTIBIAL PROSTHESIS MASS ON GAIT ASYMMETRIES Seth et al. 2020 CPOJ prosthesis mass may have on gait asymmetry and by extension health outcomes, of adults with TTA. Previous efforts to determine the relationship between prosthesis mass and gait asymmetry have been completed, but only with limited application. In one mathematical study, it was theorized that temporal asymmetry seen in the gait of adults with lower limb amputation may be minimized by the achievement of inertial symmetry between the intact and prosthetic side.15 Other experimental and mathematical studies, however, found that matching prosthetic side mass completely (100%) with the intact side, through the addition of a distal load, produced an increase in temporal asymmetry.16-19 Thus, suggesting that a prosthesis should not be as heavy as the segment it replaces, particularly if mass is distributed more distally. However, the experimental evidence is limited regarding smaller prosthesis mass increments (<100%) and the application of mass at proximal locations. This would result in a prosthesis heavier than its original weight, but lighter in comparison to the intact side. Mass alterations at proximal locations are common clinically, subsequent to changing prosthesis components, i.e., socket, liner or socket/pylon interface adapters. Hence, it is important to understand if these types of common increases in proximal prosthetic mass effect temporal asymmetry during walking. Further, the relationship between prosthesis mass and limb loading in people with TTA during walking has undergone limited exploration. Using a single subject design Hillary et al. examined the effects of increasing prosthesis mass by 0.53kg and 1.46kg on limb loading in an adult with TTA.20 The increase in prosthesis mass resulted in higher limb loading forces (1st vertical ground reaction force peak) on both the prosthetic and intact side.20 These increased impact forces are concerning as they may have consequences for bone and joint health.8,9 Increasing prosthesis mass, through proximal load application, will make a prosthesis heavier, but may not always make the prosthetic side as heavy as the intact side (or 100% matching). Currently, it is unclear whether such mass increments to a prosthesis, i.e., less than 100% matching, will influence gait temporal and loading asymmetry of individuals with TTA. Further, the majority of current evidence on prosthesis mass is based on short acclimation periods prior to experimental data collection, which may yield differing results. The purpose of this study is to explore the effects of increasing prosthesis mass on temporal and limb loading asymmetry in people with TTA following seven days of acclimation and community use. METHODOLOGY A counterbalanced repeated measures study was conducted to examine the temporal and limb loading asymmetry in adults with TTA under three load conditions: original prosthesis mass with no load added (No load; NL), original prosthesis mass with 30% of the mass difference between prosthetic and intact side added (Light load; LL), and original prosthesis mass with 50% of the mass difference between prosthetic and intact side added (Heavy load; HL). The choice of the mass conditions was based on a pilot unpublished retrospective analysis of 12 medical charts of individuals with TTA. On an average, 30% (450g) and 50% (750g) of the mass difference between the prosthetic and intact side may represent the approximate mass change that may happen as a result of changing common prosthesis components. Participants Participants were included if they were male or female, between the ages of 18 to 70 years, had a unilateral amputation ≥ one year prior to participation and were community ambulators. Further, to minimize potential effects due to recent changes in prosthesis or any components, only individuals using their current prosthesis for at least three months were included. Potential participants were excluded if they had a health condition (cardiac, pulmonary, or musculoskeletal) limiting their ability to walk or if their prosthetic side was heavier than their intact side. Based on previous literature on this topic,16,17,19 the study was planned to enroll a minimum of 6 individuals with TTA. Informed written consent was obtained before participation. The study was approved by Stony Brook University’s Institutional Review Board. Limb mass properties Body mass (kg), height (m), residual limb and prosthesis measurements were ascertained. Prosthetic side mass (PSM) was calculated as the sum of prosthesis mass (with liner and shoe on a scale) and residual limb mass. To calculate residual limb mass, the residual limb was mathematically modeled as the frustum of a right circular cone using circumferential and length measurements as inputs and assuming a uniform tissue density of 1.1gm/cm3.21 In order to calculate intact side mass, body mass (BM) needed to be adjusted to account for the amputated limb segment17: Equation 1: 𝐴𝐵𝑀 = 𝐵𝑀− 𝑃𝑆𝑀 (1−𝑐) Where, BM is measured body mass while wearing the prosthesis, PSM is prosthetic side mass, and c is the amount of mass accounted by an intact shank and foot (0.057 for men and 0.061 for women).17 Then, using ABM the intact side shank and foot mass was calculated from standard limb segment mass estimates (5.7% for men and 6.1% for women).22 In order to determine the load amount for each participant, the difference between prosthetic side and intact side mass was calculated and 30% (LL condition) and 50% (HL condition) of that difference was identified. https://doi.org/10.33137/cpoj.v3i2.34609 3 Seth M, Hou W, Goyarts L.R, Galassi J.P, Lamberg E.M. Effect of transtibial prosthesis mass on gait asymmetries. Canadian Prosthetics & Orthotics Journal. 2020;Volume 3, Issue 2, No.5. https://doi.org/10.33137/cpoj.v3i2.34609. ISSN: 2561-987X EFFECT OF TRANSTIBIAL PROSTHESIS MASS ON GAIT ASYMMETRIES Seth et al. 2020 CPOJ Research protocol Participants attended three testing sessions scheduled one week apart (Figure 1), during which the load conditions (NL, LL and HL) were examined. During session one, study purpose and procedures were explained and the mass difference between limbs was calculated. Participants were then prepared for baseline data collection with their original prosthesis mass (NL). Following NL data collection, prosthesis mass was altered by adding either 30% (LL) or 50% (HL) of the mass difference between limbs. Mass was added circumferentially to the prosthesis socket at a proximal location that was 20 to 30% of the non-prosthetic side length. Flexible car wheel balancing weights (BADA steel tape-a-weight, Hennessey Industries LLC, LaVergne, TN, USA) were affixed at the desired location with one sided tape that was further held in place by wrapping Coban (3M™, Saint Paul, MN, USA) to prevent any loosening or slippage. The load sequence was altered between participants such that half received HL then LL and half received LL then HL. Once the mass was altered participants went home to use their prosthesis for seven days during their daily activities. Figure 1: Timeline of research sessions for one participant. The order of HL or LL was counterbalanced so that half received HL at session 1 then LL at session 2 and half received LL at session 1 then HL at session 2. NL= No load, LL= Light load, HL= Heavy load. Data collection For all sessions, participants were prepared with the application of passive reflective markers over standardized anatomical landmarks on their pelvis and lower limbs using the modified Hellen Hayes marker set-up.23 On the prosthesis, markers were placed at locations corresponding to the intact side. Participants walked at their comfortable walking speed over a 10m walkway, while kinematic data was captured using a ten camera Vicon Motion System® (Oxford Metrics, Oxford, UK), recording at 100 Hz. In addition, ground reaction force (GRF) was captured through AMTI® force-plates (Watertown, USA) recording at 1000Hz. Data analysis Customized scripts developed in Matlab® (Mathworks, Inc.), were used to analyze and process the temporal and GRF data. Gait events (heel strike and toe off) were identified from the raw marker data extracted from Nexus VICON using the methodology proposed by Zeni et al.24 Temporal measures of stance time (heel strike to toe off) and swing time (toe off to heel strike) were determined for each leg. GRF data was processed using a zero-lag low-pass 4th order Butterworth filter, with a cutoff frequency of 20Hz. The beginning and end of stance phase for the vertical component of GRF (vGRF) was established using a vGRF threshold of 20N.25 The vGRF was normalized to each participant’s body mass (with prosthesis) and subsequently the maximum value during the 1st half of the stance phase was identified (vGRF Peak) and the area under the curve throughout the entire stance phase was calculated (vGRF Impulse). Symmetry indices (SI) for temporal and loading measures were calculated for each load condition26 where, I and P refer to intact and prosthetic side values. The SI ranges between 0 and 1, where 0 represents perfect symmetry. A positive SI indicates a higher value for the intact side: Equation 2: 𝑆𝐼 = (𝐼−𝑃) (𝐼+𝑃)∗0.5 Statistics A single-factor repeated measures analysis of variance (ANOVA) was used to test for differences in SI for stance time, swing time, vGRF Peak and vGRF Impulse among the NL, LL, and HL conditions. An alpha level of 0.05 was used to evaluate significance. RESULTS Study Sample Eight male individuals with unilateral TTA who ambulate without assistive devices participated in this study (Table 1). Overall, 19 inquiries were received, and 14 individuals were screened over the phone. Of these 14, nine consented while five did not meet inclusion criteria or stated they did not want to add load to their device. Of the nine consented, one was subsequently excluded because the prosthetic side mass was higher than the intact side mass. Table 2 presents the mean of the calculated mass values. SESSION 1 1) Informed consent 2) Limb mass difference calculated 3) Baseline data collection with NL 4) Prosthesis mass altered to HL SESSION 2 1) Data collection with HL 2) Prosthesis mass altered to LL SESSION 3 1) Data collection with LL 2) Prosthesis mass returned to NL 1-week acclimation 1-week acclimation https://doi.org/10.33137/cpoj.v3i2.34609 4 Seth M, Hou W, Goyarts L.R, Galassi J.P, Lamberg E.M. Effect of transtibial prosthesis mass on gait asymmetries. Canadian Prosthetics & Orthotics Journal. 2020;Volume 3, Issue 2, No.5. https://doi.org/10.33137/cpoj.v3i2.34609. ISSN: 2561-987X EFFECT OF TRANSTIBIAL PROSTHESIS MASS ON GAIT ASYMMETRIES Seth et al. 2020 CPOJ Table 1: Participant demographics and clinical characteristics. Participant Age (yrs) Body mass (kg)* Height (m) Time since amputation (yrs) Amputation etiology Side 01 48 85.0 1.88 6 Infection Left 02 29 101.0 1.72 4 Neurofibromatosis Left 03 50 113.1 1.86 1 Trauma Left 04 62 117.5 1.83 2 Vascular Right 05 26 95.0 1.84 14 Cancer Left 06 57 91.6 1.81 7 Trauma Left 07 30 75.0 1.79 30 Congenital Right 08 59 88.5 1.76 59 Congenital Right Mean (SD) 45.1 (14.7) 95.8 (14.2) 1.81 (0.05) 15.4 (20) - - * Body mass was measured with the prosthesis and shoe on the participant. Table 2: Estimated mass properties of the prosthetic and intact side. Measure Mean (SD) Range Adjusted body mass 96.9 (14.8) 75.4 – 119.5 Non-prosthetic side mass 5.5 (0.8) 5.3 – 6.8 Prosthetic side mass 4.5 (0.4) 3.9 – 5.2 Mass difference between sides 1.0 (0.7) 0.4 – 2.3 Load conditions Light Load (30% of difference) 0.3 (0.2) 0.1 – 0.7 Heavy Load (50% of difference) 0.5 (0.3) 0.2 – 1.1 All values in kg. Non-prosthetic side mass refers to estimated shank and foot mass of non-prosthetic side. Prosthetic side mass refers to the sum of estimated residual limb mass and prosthesis mass (with shoe, liner and suspension system). Main results The mean (SD) for stance time, swing time, vGRF Peak and vGRF Impulse on the prosthetic and intact side with the three load conditions (NL, LL and HL) and the SI are presented in Table 3. The mean (SD) stance time SI and swing time SI for the three load conditions is also presented in Figure 2. The mean vGRF profile of the participants for the prosthetic and intact side with the three load conditions (NL, LL and HL) is presented in Figure 3. Following seven days of acclimation, increasing the mass of the prosthesis by 30% and 50% did not significantly affect the SI values between the prosthetic and intact side for temporal (Stance time: p=0.61; swing time: p=0.13) and loading measures (vGRF Peak: p=0.95; vGRF Impulse: p=0.55). Figure 2: Mean stance time SI (A) and swing time SI (B) across the three load conditions. Error bars represent 1 SD. Negative SI values indicate that intact side values were higher. NL= No load, LL= Light load, HL= Heavy load. Table 3: Mean (SD) temporal and GRF measures for the three load conditions. NL= No load, LL= Light load, HL= Heavy load. NL LL HL Sig. (p-value) P NP SI% P NP SI% P NP SI% Stance time (secs) 0.77 (0.04) 0.80 (0.05) 3.49% (3.79) 0.79 (0.04) 0.80 (0.05) 2.15% (2.77) 0.76 (0.05) 0.78 (0.05) 2.69% (2.23) 0.61 Swing time (secs) 0.40 (0.03) 0.39 (0.02) -3.32% (5.23) 0.41 (0.01) 0.39 (0.02) -4.46% (5.16) 0.40 (0.02) 0.39 (0.02) -1.91% (5.86) 0.13 vGRF Peak (BM) 1.03 (0.07) 1.14 (0.12) 9.70% (10.70) 1.05 (0.07) 1.15 (0.07) 9.06% (6.20) 1.06 (0.08) 1.16 (0.09) 9.51% (9.35) 0.95 vGRF Impulse (BM) 0.56 (0.04) 0.60 (0.04) 7.99% (7.11) 0.57 (0.03) 0.61 (0.04) 6.89% (6.32) 0.56 (0.04) 0.60 (0.03) 6.73% (7.53) 0.55 P: mean (SD) raw score for the prosthetic side; NP: mean (SD) raw score for the non-prosthetic side; SI%: symmetry index % between the prosthetic and non- prosthetic side; BM: normalized to body mass. The p-values between the three load conditions are for the SI% 3.49% 2.15% 2.69% 0% 2% 4% 6% 8% NL LL HL Stance time SIA -3.32% -4.46% -1.91% -11% -7% -3% NL LL HL Swing time SI B https://doi.org/10.33137/cpoj.v3i2.34609 5 Seth M, Hou W, Goyarts L.R, Galassi J.P, Lamberg E.M. Effect of transtibial prosthesis mass on gait asymmetries. Canadian Prosthetics & Orthotics Journal. 2020;Volume 3, Issue 2, No.5. https://doi.org/10.33137/cpoj.v3i2.34609. ISSN: 2561-987X EFFECT OF TRANSTIBIAL PROSTHESIS MASS ON GAIT ASYMMETRIES Seth et al. 2020 CPOJ Figure 3: Mean normalized vGRF activity of the prosthetic (A) and intact side (B) across the three load conditions (NL, LL and HL). The shaded region represents 1±SD of the group mean27 for that load condition. NL= No load, LL= Light load, HL= Heavy load. DISCUSSION This study explored the impact of increasing prosthesis mass on temporal and limb loading asymmetry in people with TTA following seven days of acclimation and community use. According to the findings, increasing prosthesis mass by either 30% or 50% of the mass difference between legs did not significantly alter temporal and limb loading asymmetry. These findings imply that a range of prosthesis mass may exist for this population. This ultimately may allow clinicians flexibility when choosing prosthesis components knowing that increasing mass within this range may not negatively impact gait asymmetries. Temporal asymmetry Literature on transtibial prosthesis mass has largely focused on examining temporal asymmetry during walking.15-19 Perhaps the proposed association between limb inertia and swing time or the secondary complications as a result of an asymmetrical gait, necessitate the examination of temporal measures following mass alterations. In this study, stance and swing time asymmetries were not significantly affected as a result of the mass alteration. Previous studies examining stance and swing time asymmetries have found that matching prosthetic side mass to the intact side (100% load condition) increases asymmetry.16-19 The same however, is not true for partial matching (50% load condition). Similar to the findings of the current study, Smith and Martin report no effect of the 50% load condition on stance and swing time asymmetry.17 However, Mattes et al. observed a significant increase in temporal asymmetry with the 50% load condition.16 Hence, Mattes et al., observed a linear response to the mass increase; as prosthesis mass increased, stance time on the intact side and swing time on the prosthetic side increased, thereby increasing limb temporal asymmetry.16 Interestingly, Mattes et al., used a similar approach as the current study to estimate mass difference between limbs. However, they achieved the 50% condition by adding an average load of 0.85kgs,16 while this study used an average of 0.54kgs, which may partially explain the divergence in findings. Perhaps the key difference between the studies is the length of acclimation period used, ten minutes by Mattes et al.,16 versus one week in this study. The longer acclimation time provided in this study may have allowed participants to stabilize their gait, leading to the lack of statistically significant findings. Although, in previous research it has been observed that stance and swing time asymmetries (as a result of the 100% load condition) persisted even after eight days of acclimation19; this finding, doesn’t necessarily rule out the importance of acclimation. It is more likely that individuals with TTA are not well suited to carry a prosthesis that weighs as much as the amputated segment. Limb loading asymmetry Limb loading asymmetry (measured using vGRF Peak and vGRF Impulse SI), was not affected as a result of the mass increase, i.e., no differences were observed between the three load conditions (NL, LL and HL). The positive SI values (Table 3) indicate that participants in this study loaded their intact side more than the prosthetic side, consistent with previous findings in the literature.5-8 The addition of load, however, did not significantly change the proportion of loading experienced by the intact side. Prosthetic side % Gait cycle Intact side N o rm a li z e d v G R F N o rm a li z e d v G R F % Gait cycle https://doi.org/10.33137/cpoj.v3i2.34609 6 Seth M, Hou W, Goyarts L.R, Galassi J.P, Lamberg E.M. Effect of transtibial prosthesis mass on gait asymmetries. Canadian Prosthetics & Orthotics Journal. 2020;Volume 3, Issue 2, No.5. https://doi.org/10.33137/cpoj.v3i2.34609. ISSN: 2561-987X EFFECT OF TRANSTIBIAL PROSTHESIS MASS ON GAIT ASYMMETRIES Seth et al. 2020 CPOJ In the single subject case study by Hillery et al., the authors observed an increase in loading (vGRF Peak) on the prosthetic and the intact side with an increase in prosthesis mass.20 Based on the SI equation,26 it appears that the study participant’s loading asymmetry increased from 5.4% at baseline to 12% with the intermediate load (0.5kgs) and 21% with the heaviest load (1.5kgs).20 In contrast, vGRF Peak SI in the current study changed from an average of 9.70% at baseline to 9.1% with LL (0.3kgs) and 9.5% with the heaviest load condition, HL (0.5kg). This difference in findings may be due to the heavier loads used by Hillery et al., or perhaps the longer acclimation period given in the current study allowed users to stabilize their gait before data was collected. However, given the single-subject study design by Hillery et al., their findings may not be generalized to a larger group.20 In a more recent study of a group of active adults with TTA (n=10), Alcantara et al. examined the impact of increasing the mass of a running-specific prosthesis and biological foot on participants’ limb loading asymmetry during running.28 Prosthesis and biological foot mass were increased by adding 100g or 300g at the toe region. The mass increase, however, did not result in any significant change to limb loading asymmetry (vGRF peak SI or average-stance vGRF SI) as compared to baseline, regardless of added load (100g or 300g) or whether mass was added to the prosthesis or both limbs.28 Obvious differences exist between Alcantara et al., and the current study, such as the location of mass addition or their use of a running task. However, collectively, these findings suggest that small increases in prosthesis mass may not contribute to the limb loading asymmetry and by extension lower health outcomes generally associated with asymmetrical limb loading, such as knee or hip osteoarthritis of the intact side. The need to change a prosthesis component or replace a prosthesis completely may arise due to various personal and/or clinical reasons. Such changes may increase or decrease the overall prosthesis mass. For example, the replacement of prefabricated transtibial gel liners, which range from 202g to 722g, may alter prosthesis mass.29 Findings from this study suggest that the mass increments similar to those observed after replacing proximally located prosthesis components (such as, the socket, liners, or socket/pylon interface adapters), may not increase gait asymmetry. These findings are clinically relevant as they may serve as a guide for clinicians when routinely adjusting prosthesis mass. However, to further identify the impact of prosthesis mass on gait asymmetry, it is vital that future large-scale studies explore the impact of various commercially available prosthesis components and related functional and clinical outcomes. Moreover, future studies may consider controlling other factors that may influence gait characteristics of an individual, for example, residual limb length, quality and type of suspension between the prosthesis and residual limb, or exertion of the individual during a walking task. Study limitations The current study was limited both by a small sample size (n=8), and a male-only participant group. It is possible that the small sample size may not have had sufficient power to detect a significant change. Moreover, in order to make stronger conclusions on prosthesis mass, it remains necessary to evaluate both men and women with TTA. As part of the inclusion criteria, only individuals that were capable of independent ambulation were included within the study. Therefore, the current findings may not extend to individuals with TTA who exhibit lower levels of activity. Further, this study chose to limit mass increase to 50% of the mass difference between both limbs. To establish the upper limit of a TTP mass it may be necessary to evaluate heavier loads. Lastly, certain covariates, such as, residual limb length, quality and type of suspension between the prosthesis and residual limb or prosthesis componentry may have impacted both gait asymmetry and prosthesis mass of the study participants but were not controlled in this study. CONCLUSION Temporal and loading asymmetries during walking in people with TTA may lead to secondary complications of the intact side and increase risk of falls. Currently, it is unclear if prosthesis mass alterations that may occur during routine clinical visits will impact these gait asymmetries. This exploratory study observed that following seven days of use increases in proximally distributed prosthesis mass (up to 50% of the mass difference between legs), did not further increase temporal and limb loading asymmetry of individuals with TTA, thereby allowing more flexibility to the clinician for component selection. Confirmation of findings with large-scale studies may help identify the impact of prosthesis mass on gait asymmetry. ACKNOWLEDGEMENTS We would like to acknowledge study participants for their contribution towards this research study. DECLARATION OF CONFLICTING INTERESTS The authors have no conflicts of interests to declare. AUTHOR CONTRIBUTION Mayank Seth: Study conceptualization, analysis, data collection, data interpretation and writing - original draft Wei Hou: Assisted in study design, led statistical analysis, writing - review & editing https://doi.org/10.33137/cpoj.v3i2.34609 7 Seth M, Hou W, Goyarts L.R, Galassi J.P, Lamberg E.M. Effect of transtibial prosthesis mass on gait asymmetries. Canadian Prosthetics & Orthotics Journal. 2020;Volume 3, Issue 2, No.5. https://doi.org/10.33137/cpoj.v3i2.34609. ISSN: 2561-987X EFFECT OF TRANSTIBIAL PROSTHESIS MASS ON GAIT ASYMMETRIES Seth et al. 2020 CPOJ Laura R Goyarts: Assisted in data collection, writing - review & editing James P Galassi: Assisted in data collection, writing - review & editing Eric M Lamberg: Study conceptualization, project administration, supervision, data interpretation, writing – review & editing SOURCES OF SUPPORT This research did not receive any specific grant from funding agencies in the public, commercial or not-for-profit sectors. ETHICAL APPROVAL The study was approved by Stony Brook University’s Institutional Review Board. REFERENCES 1.Dingwell JB, Davis BL, Frazier DM. Use of an instrumented treadmill for real-time gait symmetry evaluation and feedback in normal and trans-tibial amputee subjects. Prosthet Orthot Int. 1996;20(2):101-10. DOI: 10.3109/03093649609164426 2.Nolan L, Lees A. The functional demands on the intact limb during walking for active transfemoral and transtibial amputees. Prosthet Orthot Int. 2000;24(2):117-25. DOI: 10.1080/03093640008726534 3.Nolan L, Wit A, Dudzinski K, Lees A, Lake M, Wychowanski M. Adjustments in gait symmetry with walking speed in trans-femoral and trans-tibial amputees. Gait Posture. 2003;17(2):142-51. DOI: 10.1016/s0966-6362(02)00066-8 4.Sanderson DJ, Martin PE. Lower extremity kinematic and kinetic adaptations in unilateral below-knee amputees during walking. Gait Posture. 1997;6(2):126-36. DOI:10.1016/S0966-6362(97)01112-0 5.Cutti AG, Verni G, Migliore GL, Amoresano A, Raggi M. Reference values for gait temporal and loading symmetry of lower- limb amputees can help in refocusing rehabilitation targets. J Neuroeng Rehabil. 2018;15(1):61. DOI: 10.1186/s12984-018- 0403-x 6.Engsberg J, Lee A, Tedford K, Harder J. Normative ground reaction force data for able-bodied and trans-tibial amputee children during running. Prosthet Orthot Int. 1993;17(2):83-9. DOI: 10.3109/03093649309164361 7.Powers CM, Torburn L, Perry J, Ayyappa E. Influence of prosthetic foot design on sound limb loading in adults with unilateral below-knee amputations. Arch Phys Med Rehabil. 1994;75(7):825- 9. DOI: 10.1016/0003-9993(94)90146-5 8.Gailey R, Allen K, Castles J, Kucharik J, Roeder M. Review of secondary physical conditions associated with lower-limb amputation and long-term prosthesis use. J. Rehabil. Res. Dev. 2008;45(1). DOI: 10.1682/JRRD.2006.11.0147 9.Kulkarni J, Adams J, Thomas E, Silman A. Association between amputation, arthritis and osteopenia in British male war veterans with major lower limb amputations. Clin Rehabil. 1998;12(4):348- 53. DOI: 10.1191/026921598672393611 10.Esposito ER, Wilken JM. The relationship between pelvis–trunk coordination and low back pain in individuals with transfemoral amputations. Gait posture. 2014;40(4):640-6. DOI: 10.1016/ j.gaitpost.2014.07.019 11.Vanicek N, Strike S, McNaughton L, Polman R. Gait patterns in transtibial amputee fallers vs. non-fallers: Biomechanical differences during level walking. Gait Posture. 2009;29(3):415-20. DOI: 10.1016/j.gaitpost.2008.10.062 12.Handžić I, Reed KB. Perception of gait patterns that deviate from normal and symmetric biped locomotion. Front. Psychol. 2015;6:199. DOI: 10.3389/fpsyg.2015.00199 13.Adamczyk PG, Kuo AD. Mechanisms of Gait Asymmetry Due to Push-Off Deficiency in Unilateral Amputees. IEEE Trans Neural Syst Rehabil Eng. 2015;23(5):776-85. DOI: 10.1109/TNSRE. 2014.2356722 14.Silver-Thorn B, Steege JW, Childress DS. A review of prosthetic interface stress investigations. J. Rehabil. Res. Dev. 1996. 15.Mena D, Mansour J, Simon S. Analysis and synthesis of human swing leg motion during gait and its clinical applications. J Biomech. 1981;14(12):823-32DOI:10.1016/0021-9290(81)90010-5 16.Mattes SJ, Martin PE, Royer TD. Walking symmetry and energy cost in persons with unilateral transtibial amputations: matching prosthetic and intact limb inertial properties. Arch Phys Med Rehabil. 2000;81(5):561-8. DOI: 10.1016/S0003-9993(00)90035-2 17.Smith JD, Martin PE. Effects of prosthetic mass distribution on metabolic costs and walking symmetry. J Appl Biomech. 2013;29(3):317-28. DOI: 10.1123/jab.29.3.317 18.Selles RW, Korteland S, Van Soest A, Bussmann JB, Stam HJ. Lower-leg inertial properties in transtibial amputees and control subjects and their influence on the swing phase during gait. Arch Phys Med Rehabil. 2003;84(4):569-77. DOI: 10.1053/apmr. 2003.50037 19.Smith JD, Martin PE. Short and longer term changes in amputee walking patterns due to increased prosthesis inertia. J Prosthet Orthot. 2011;23(3):114-23. DOI: 10.1097/JPO.0b013e3182248d90 20.Hillery S, Wallace E, McIlhagger R, Watson P. The effect of changing the inertia of a trans-tibial dynamic elastic response prosthesis on the kinematics and ground reaction force patterns. Prosthet Orthot Int. 1997;21(2):114-23. DOI: 10.3109/ 03093649709164539 21.Hanavan EL. A Mathematical Model of the Human Body. AMRL- TR-64-102, AD-608-463. Ohio: Aerospace Medical Research Laboratory, Wright-Patterson Air Force Base. 1964. 22.De Leva P. Adjustments to Zatsiorsky-Seluyanov's segment inertia parameters. J Biomech. 1996;29(9):1223-30. DOI:10.1016/ 0021-9290(95)00178-6 23.Kadaba MP, Ramakrishnan HK, Wootten ME. Measurement of lower extremity kinematics during level walking. J Orthop Res. 1990;8(3):383-92. DOI: 10.1002/jor.1100080310 24.Zeni Jr J, Richards J, Higginson J. Two simple methods for determining gait events during treadmill and overground walking using kinematic data. Gait posture. 2008;27(4):710-4. DOI: 10.1016/j.gaitpost.2007.07.007 25.Rueda FM, Diego IMA, Sánchez AM, Tejada MC, Montero FMR, Page JCM. Knee and hip internal moments and upper-body kinematics in the frontal plane in unilateral transtibial amputees. Gait Posture. 2013;37(3):436-9. DOI: 10.1016/j.gaitpost.2012. 08.019 https://doi.org/10.33137/cpoj.v3i2.34609 8 Seth M, Hou W, Goyarts L.R, Galassi J.P, Lamberg E.M. Effect of transtibial prosthesis mass on gait asymmetries. Canadian Prosthetics & Orthotics Journal. 2020;Volume 3, Issue 2, No.5. https://doi.org/10.33137/cpoj.v3i2.34609. ISSN: 2561-987X EFFECT OF TRANSTIBIAL PROSTHESIS MASS ON GAIT ASYMMETRIES Seth et al. 2020 CPOJ 26.Robinson RO, Herzog W, Nigg BM. Use of force platform variables to quantify the effects of chiropractic manipulation on gait symmetry. J Manipulative Physiol Ther. 1987;10(4):172-6. 27.Musall S. stdshade MATLAB Central File Exchange 2010 Available from: https://www.mathworks.com/matlabcentral/ fileexchange/29534-stdshade. 28.Alcantara RS, Beck ON, Grabowski AM. Added lower limb mass does not affect biomechanical asymmetry but increases metabolic power in runners with a unilateral transtibial amputation. Eur J Appl Physiol. 2020;120(6):1449-56. DOI: 10.1007/s00421-020-04367-9 29. Selinger D. Weight of Prefabricated Transtibial Gel Prosthetic Liners A Standardized Sample. Resident Research Project. California State University, Dominguez Hills. 2004. https://doi.org/10.33137/cpoj.v3i2.34609