cana-3610	Sharmila Vallem	Design of a High speed Low Low-Power Latched Comparator for Medical Implants	2025	18	.pdf	application/pdf	6921	390	49	DTDLC Recommended comparator Power utilization (nW) Pre- Layout 31.45 59.87 31.29 Post - Layout 34.76 67.28 32.97 Delay (ns) Pre- Layout 38.64 42.54 31.86 Post - Layout 42.76 49.44 36.54 Area 3.63μm X 4.1μm 3.7μm X 4.2μm 3.6μm X 4.12μm Figure 17: Layout of proposed comparator As technological advances shrinks and fabrication complexity increases, it is essential to replicate comparators across different process corners. Latch comparators have gained increased popularity over static comparators within low-power medical applications.	cache/cana-3610.pdf	txt/cana-3610.txt