id	author	title	date	pages	extension	mime	words	sentence	flesch	summary	cache	txt
eurjchem-721	Salama, Ismail; Gomaa, Mohamed Sayed	Comparative determination of miconazole, nystatin, hydrocortisone and neomycin by HPTLC/HPLC-DAD	2013	6	.pdf	application/pdf	4727	272	59	Excellent linearity was observed for all of the standard calibration curves, and the correlation coefficients were above 0.9997. HPTLC limits of quantitation were 15.37×10‐2, 9.8×10‐2, 13.32×10‐2 and 15.19×10‐2 µg/mL for MIC, NYS, HCA and NS respectively, whereas HPLC limits were 6.80×10‐2, 6.56×10‐2, 1×10‐1 and 10.16×10‐2 µg/band for MIC, NYS, HCA and NS respectively. Furthermore, the main objectives of this work were to develop, validate and compare two new HPLC‐DAD and HPTLC methods for the simultaneous determination of NYS, MIC, HCA and NS in pharmaceutical preparations.	cache/eurjchem-721.pdf	txt/eurjchem-721.txt