Nova Biotechnol Chim (2018) 17(1): 1-15 DOI: 10.2478/nbec-2018-0001  Corresponding author: svetlana.hrouzkova@stuba.sk Nova Biotechnologica et Chimica Review Present state and applications of single drop microextraction for the determination of harmful organic compounds and pollutants Silvia Zichová, Adriana Brisudová and Svetlana Hrouzková Institute of Analytical Chemistry, Faculty of Chemical and Food Technology, Slovak University of Technology in Bratislava, Radlinského 9, Bratislava, SK-812 37, Slovak Republic Article info Article history: Received: 29th March 2018 Accepted: 20th June 2018 Keywords: Harmful organic compounds Microextraction Pollutants Single drop microextraction Abstract Single drop microextraction (SDME) nowadays earns an increasing attention by scientists due to its simplicity, low cost and the need for only common laboratory equipment. This microextraction technique combines sample cleanup and pre-concentration of analytes in one step. Furthermore, a significant reduction in the amount of organic solvents needed comparing to standard LLE techniques places SDME into the position of environmental friendly extraction techniques. SDME is a straightforward technique in which a micro-drop of solvent is suspended from the tip of a conventional micro-syringe and then it is in a direct contact with a sample solution in which it is immiscible or it could be suspended in the headspace above the sample. The paper overviews developments of the state- of-the-art SDME techniques for the extraction of harmful organic compound and pollutants from environmental, food and biological matrices. Key extraction parameters essential for SDME performance were described and discussed.  University of SS. Cyril and Methodius in Trnava Introduction The toxicity of pesticides and pollutants and their harmful environmental effects are attracting the attention of the society and interest in the identification and quantification of such compounds in various matrices have aroused. Thus, the importance is put on the development of the faster and selective analytical methodology, suitable for the determination of the low levels of these compounds. The limits of quantitation (LOQs) of pesticides and pollutants are usually in mg/L or below. Therefore, in order to accomplish the quantitation of these compounds in various samples, a pre-treatment step, extraction and pre-concentration step are required. Generally, when conventional liquid phase extraction is used, the preconcentration of the analytes is very small or not even obtained. A lot of research efforts have been focused on the development of combined sample preparation techniques, when conventional liquid phase extraction is in the combination with a proper liquid phase microextraction technique or sorptive based microextraction technique, in order to achieve additional preconcentration of the analytes. Contrary to this, in the case of the single drop microextraction (SDME) these two processes are performed in a single step (De Souza Pinheiro et al. 2011; Andraščíková et al. 2016). Currently SDME has been increasingly used in the analysis of pesticides and other environmental Bereitgestellt von Slovenská poľnohospodárska knižnica | Heruntergeladen 28.02.20 08:24 UTC Nova Biotechnol Chim (2018) 17(1): 1-15 2 pollutants in various matrices because it is simple, cheap, fast, effective and consuming only microliters of organic solvents per analysis. SDME technique is based on the principle of a distribution of analytes between a microdrop of the organic solvent and an aqueous phase or headspace phase above the solid/liquid sample. Using volume of sample in a milliliter scale and microliters of organic solvent leads to a high enrichment factor obtained. The SDME procedure using conventional microsyringe has the advantage of combining manipulation and injection of the sample into a single-step extraction and thus simplify the extraction process. However, some problems such as microdrop instability leading to the dislodgement of the drop or dislocation of the drop from the needle tip, and partial solvent loss caused by the high temperature and high solubility of the drop in the sample during extraction may reduce the repeatability of the method (Jeannot et al. 2010; Ramos 2012). Sampling repeatability requires the use of the proper syringe needle for the extraction. The use of the microdrop as a collector phase of analytes from the gaseous samples (Liu and Dasgupta 1995) and liquid samples (Liu and Dasgupta 1996) was described for the first time in the nineties of the 20th century. By the time, SDME has undergone many technical changes, and it is known in various modes. Seven different modes of solvent microextraction that belong to the category of single drop microextraction are currently in use for various applications. The base classification of SDME modes is dividing into two sub-categories, two and three-phase techniques, which exist in equilibrium (Table 1). All modes are based on the principle of passing the analytes from the sample directly, or through the other phase (e.g. headspace) or mediator to the final extraction solvent having a volume in microliters (Ramos 2012). There are more advances to classify SDME, e.g. according to hydrodynamic features, static and dynamic modes of operation are distinguished (Alexovič et al. 2016). Use of the extraction solvent, which is volatile, e.g. hexane or toluene, makes the application directly compatible with gas chromatography (Jeannot et al. 2010). Moreover, combinations of SDME with high performance liquid chromatography (HPLC) and capillary electrophoresis were reported (Amde et al. 2015; García-Vázquez et al. 2016; Yohannes et al. 2016). Table 1. Overview of the SDME modes. SDME mode Acronym Two-phases techniques direct immersion DI-SDME continuous flow microextraction CFME drop-to-drop microextraction DDME directly suspended droplet microextraction DSDME Three-phases techniques headspace HS-SDME liquid-liquid-liquid microextraction LLLME solvent-supported microextraction SSME SDME has become frequently used for the extraction of a broad scope of compounds for numerous analytical application due to the advantages as reviewed by more papers (Lamboropoulou et al. 2007; Jeannot et al. 2010; Jain and Verma 2011). In this review, the focus was given to the three main application areas of SDME for the extraction of the harmful organic compounds and pollutants in environmental, food and biological samples in the recent years. Environmental samples SDME has been used to extract a wide range of organic compounds from the environmental samples, mostly several types of water and soil. Overview of the applicability of SDME for environmental sample analysis is shown in the Table 2. SDME was an appropriate extraction technique for liquid samples such as tap water (Kaykhaii et al. 2005; Lamboropoulou et al. 2007; Yohannes et al. 2016), river water (Kaykhaii et al. 2005; Saraji and Bankhshi 2005; Ahmadi et al. 2006; De Souza Pinheiro et al. 2009; De Souza Pinheiro et al. 2011; Soares et al. 2014; Amde et al. 2015; Yohannes et al. 2016), lake water (Wu et al. 2008; Xie et al. 2014; Amde et al. 2015; Yohannes et al. 2016), surface water (Lopez- Blanco et al. 2003; Lamboropoulou et al. 2007; Santos et al. 2017), drinking water (Lopez-Blanco et al. 2003; Ahmadi et al. 2006; Carlos et al. 2013), effluent and influent water (Amde et al. 2015), ground water (Yohannes et al. 2016; Santos et al. 2017), water from the farm (Ahmadi et al. 2006), Bereitgestellt von Slovenská poľnohospodárska knižnica | Heruntergeladen 28.02.20 08:24 UTC Nova Biotechnol Chim (2018) 17(1): 1-15 3 sea water (Tian et al. 2014) and other water samples (Przyjazny and Kokosa 2002). Furthermore, SDME was applied for the extraction of the pesticides from the soil samples (Salemi et al. 2013; Williams et al. 2014; Soares et al. 2015). Organic pollutants, such as pesticides (organochlorine pesticides (OCPs), organophosphorus pesticides (OPPs)), fungicides, aliphatic amines, volatile organic compounds, phenols, polyaromatic hydrocarbons (PAHs), nitro- PAHs and quinones extracted by SDME were mostly reported. Generally, two SDME modes were employed, DI and HS, nevertheless DI was preferably used due to the chemical properties of analyte and the type of the sample. Special pre-treatment was not necessary for the majority of selected applications, except addition of NaCl to the sample for increasing the ionic strength by Lamboropoulou et al. (2007), Wu et al. (2007) and Yohannes et al. (2016), adjustment of pH to 2 – 6 reported by Saraji and Bankhshi (2005), and filtration in the case of surface and ground water (Santos et al. 2017). The volume of a sample varied in the range of 1 mL – 30 mL, the most commonly selected volume of the sample was 5 mL. The pre-treatment of the solid samples was more difficult in comparison with water samples. Drying, sieving, ultrasonic extraction, sonication, centrifugation and final dilution with an appropriate solvent were used due to the matrix of the sample. The amount of sample was ranging from 2 g to 5 g (Salemi et al. 2013; Williams et al. 2014; Soares et al. 2015). The selection of a suitable extraction solvent is a crucial step for SDME performance. According to the extracted analytes, mostly organic solvents differing in Kow compatible with GC and HPLC were chosen, in order to obtain the highest extraction efficiency. Toluene, provided stable drop and avoided bubble formation during extraction according to vapour pressure lower than other solvents, was mostly used in direct immersion (DI) mode (Lamboropoulou et al. 2007; De Souza Pinheiro et al. 2009; De Souza Pinheiro et al. 2011; Tian et al. 2014; Santos et al. 2017). However, the bubble formation in the drop was required for the extraction of pesticides from lake water by Xie et al. (2014) by exposing the liquid-gas pendant drop (CPD) to the sample. The stability of the CPD was investigated by monitoring the formation of 1 μL organic droplets (toluene, chlorobenzene) containing different volumes of air bubble in a sample solution resulting in the better stability of the drop using chlorobenzene. A formation of a liquid-gas CPD was performed using a novel extraction assembly consisting of a 10 mm long quartz capillary, a funnel-like PMMA cap and a 5 mL microsyringe. The limitation of droplet instability in comparison with DI-SDME was overcome, larger microdrop size and higher stirring rates could be employed, as well as liquid-gas CPD has a potential to be fully automated by using a syringe pump (Xie et al. 2014). A 70- to 135-fold enrichment of pesticides was obtained. In comparison to a conventional SDME, improvement of the extraction efficiency could be ascribed by the increased surface area of the microdrop. The bubble in drop (BID) was studied by Williams et al. (2014) when chloroform was used as the extraction solvent. For the extraction of OCPs from tap and surface water, toluene and isooctane gave similar results for the majority of target analytes, but toluene was selected due to the selectivity and no significant loss of solvent during extraction (Lamboropoulou et al. 2007). The similar observation was reported for n-hexane that was selected due to its lower solubility in water, which increases the stability of the microdrop in comparison to the use of toluene. The analyte extraction was not significantly affected by the type of extraction solvent (n-hexane or toluene) (Soares et al. 2014). n-Hexane was suitable for the extraction of OCPs and pyrethroids from drinking water (Carlos et al. 2013), multiclass pesticides from water and soil (Soares et al. 2014; Soares et al. 2015). For the extraction of the atrazine, desethyl-atrazine, desisopropyl-atrazine from the water samples, toluene, n-hexane, and cyclohexane were investigated, however, due to their low viscosity and density, the microdrop was not stable. Formation of the stable drop under the vigorous stirring and higher extraction efficiency was reached using 1-octanol as the extraction solvent (Yohannes et al. 2016). Solubility of the extraction solvent in water had a great effect for the extraction of 2 pesticides, α-endosulphane and β-endo- sulphane from water. n-Hexane didn’t exhibit a good extractability of these compounds, because Bereitgestellt von Slovenská poľnohospodárska knižnica | Heruntergeladen 28.02.20 08:24 UTC Nova Biotechnol Chim (2018) 17(1): 1-15 4 S o a re s e t a l. (2 0 1 4 ) A m d e e t a l. (2 0 1 5 ) T ia n e t a l. ( 2 0 1 4 ) X ie e t a l. ( 2 0 1 4 ) C a rl o s e t a l. (2 0 1 3 ) Y o h a n n e s e t a l. (2 0 1 6 ) Bereitgestellt von Slovenská poľnohospodárska knižnica | Heruntergeladen 28.02.20 08:24 UTC Nova Biotechnol Chim (2018) 17(1): 1-15 5 D e S o u z a P in h e ir o e t a l. (2 0 1 1 ) D e S o u z a P in h e ir o e t a l. (2 0 1 1 ) P rz y ja z n y a n d K o k o sa ( 2 0 0 2 ) K a y k h a ii e t a l. (2 0 0 5 ) A h m a d i e t a l. (2 0 0 6 ) L o p e z -B la n k o e t a l. ( 2 0 0 3 ) S a ra ji a n d B a n k h sh i (2 0 0 5 ) L a m b o ro p o u lo u e t a l. ( 2 0 0 7 ) S a n to s e t a l. (2 0 1 7 ) Bereitgestellt von Slovenská poľnohospodárska knižnica | Heruntergeladen 28.02.20 08:24 UTC Nova Biotechnol Chim (2018) 17(1): 1-15 6 W u e t a l. ( 2 0 0 7 ) S a le m i e t a l. (2 0 1 3 ) W il li a m s e t a l. (2 0 1 4 ) S o a re s e t a l. (2 0 1 5 ) Bereitgestellt von Slovenská poľnohospodárska knižnica | Heruntergeladen 28.02.20 08:24 UTC Nova Biotechnol Chim (2018) 17(1): 1-15 7 it is more soluble in water in comparison to isooctane. Isooctane, being solvent of low polarity, provided higher extractive efficiency and reproducibility for both analytes (Lopez- Blanco et al. 2003). Extraction of the phenols from the river water by DI-SDME followed by in- syringe derivatization was investigated by Saraji and Bankhshi (2004). Several organic solvents in the broad range of polarity were investigated, resulted hexyl acetate being the most suitable for the extraction. Hexyl acetate provided the selective extraction of the phenols, in comparison to other solvent, when 4­nitrophenol was not extracted. Furthermore, the manipulation with hexyl acetate microdrop was acceptable without any stability problems during the extraction, however, chromatographic peak overlaps were observed (Saraji and Bankhshi 2004). The volatile compounds were preferably extracted by using HS mode. Octanol, benzylalkohol and β-cyclodextrine were investigated as potential extraction solvents (Przyjazny and Kokosa 2002; Kaykhaii et al. 2005; Salemi et al. 2013). HS- SDME was found as an efficient technique for the extraction of the organic compounds such as benzene, toluene, ethylbenzene and xylenes from the water samples. Gas chromatography with flame ionization detector (GC-FID) was employed. For the highest extraction efficiency under the experimental conditions, solvents with a high boiling point and low vapour pressure were investigated, so as to minimize the evaporation of the microdrop during the extraction. 1-Octanol and n-hexadecane provided satisfactory extraction of the analytes. However, less impurities using n-hexadecane were interfering the analyte signal (Przyjazny and Kokosa 2002). The specific application of HS-SDME which was free of organic solvent was provided by Wu et al. (2007), when β-cyclodextrine was used as extraction solvent for PAHs from the lake water. Nanofluid (NF) prepared by dispersion of ZnO nanoparticles in 1-hexyl-3-methylimidazolium- hexafluorophosphate was reported as an extraction solvent for three fungicides in water. A syringe tip cap (obtained from 5 mL disposable syringe) was used as a holder of the microdrop instead of the syringe needle. The microdrop was immersed on the syringe tip cap by microsyringe and after the extraction, the microdrop was placed in an Eppendorf vial containing 100 μL methanol with the aim to remove nanoparticles. Nowadays, NFs earn an attention due to the frequent applications in many research directions. Replacing conventional solvent in liquid–liquid system by NPs can cause an increase of mass transfer coefficient and increase the efficiency of extraction as the main goal (Amde et al. 2015). Volume of the drop is the next important factor for the efficient extraction. In general, increasing the drop volume results in the significant improvement of the extraction efficiency. However, when the microdrop volume increases further, the drop became unstable due to the gravity. The volume of the drop was in the range of 0.9 μL (Ahmadi et al. 2006) and 10 μL (Wu et al. 2007; Amde et al. 2015). Drop volume 1 μL was the most frequently applied, what can be explain by the found consensus between the volume, stability and extraction capacity of the drop (Kaykhaii et al. 2005; De Souza Pinheiro et al. 2009; De Souza Pinheiro et al. 2011; Williams et al. 2014; Xie et al. 2014; Santos et al. 2017). Sampling repeatability of the microdrop requires the use of the proper syringe needle for the extraction. Generally, 10 μL Hamilton syringe routinely used for GC was frequently used as a holder of the microdrop during the extraction process. The needle should have a minimum dead volume, for instance using a 1 μL microsyringe, no dead volume is occurred. The modification of needle tip, e.g. cones tip, or modified microsyringe causes increasing adhesion force between needle tip and drop, thereby increasing drop stability and achieving a high stirrer speed (up to 700 rpm for the cones tip and up to 1,700 rpm for the modified microsyringe) (Ahmadi et al. 2006). Furthermore, improved holding of the microdrop and larger extraction solvent volumes used for the extraction were reached by attaching a 2-mm long cone that was cut off from a 200 μL pipette tip onto the needle tip of a GC microsyringe (Tian et al. 2014). The similar approach was used for the modification of the HPLC syringe, which was supported by a part of the pipette tip, thus the larger microdrop was obtained (Wu et al. 2007). To avoid impurities, a rinse of the microsyringe before extraction was presented, mostly by the extraction solvent (Przyjazny and Kokosa 2002; Kaykhaii et al. 2005; Ahmadi et al. 2006; Wu et al. 2007; Bereitgestellt von Slovenská poľnohospodárska knižnica | Heruntergeladen 28.02.20 08:24 UTC Nova Biotechnol Chim (2018) 17(1): 1-15 8 Salemi et al. 2013; Tian et al. 2014; Yohannes et al. 2016; Santos et al. 2017). Sample agitation is an important factor for the reduction of the extraction time. From the film theory of convective-diffusive mass transfer, the agitation of the sample solution enhances the extraction efficiency and reduce the extraction time, since the thickness of the stagnant film around the extracting phase (i.e. Nernst diffusion film) decreases with increasing stirring rate, and it results in faster extraction rate (Pena-Pereira et al. 2010). Magnetic stirring, mechanical vibration, continuous flow and syringe plunger motion are usually used for the sample agitation to increase the amount of convective mixing or interfacial contact area for the dynamic modes of SDME. The time required to reach equilibrium in SDME dependent on the type and degree of agitation, phase volumes, interfacial contact area and equilibrium distribution constant (Jeannot and Cantwell 1996; Jeannot et al. 2010). Magnetic stirring was applied for the majority of the applications, with the velocity range from 155 rpm (Soares et al. 2014) to 1,300 rpm (Ahmadi et al. 2006). High velocity of agitation was used for the HS mode, when the drop is not in the direct interaction with a sample. For the extraction of the atrazine and metachlor from soil, no stirring was used and the extraction was performed under static conditions (Williams et al. 2014). The positive effect of the ultrasound waves has been used in SDME for the OPPs extraction from the soil sample (Salemi et al. 2013). For HS-SDME, the temperature is a key parameter to speed-up the rate of mass transfer of compounds from the sample to the headspace increasing the number and the amount of compounds of interest moved to the gas phase. Increased temperatures lead to the decrease of the extractant-headspace phase distribution constant and this results in the decrease of sensitivity of the method. Cooling of the extraction solvent, while the sample could be heated, is the promising solution of this problem. However, such an arrangement of experiment brings apparatus design complications, therefore, it is useful for ultra-trace analyses or for very volatile analytes with a low possibility to transfer to the headspace at ambient conditions (Jeannot et al. 2010). Extraction temperature was in the range from 17 °C (Soares et al. 2014) to 60 °C (Salemi et al. 2013). The increase of the ionic strength of the water- based sample can affect the analyte transfer by the increase of the amount of salt to the aqueous phase (Wang et al. 2017). The majority of the publications were reported to be without salt addition. Food samples SDME extraction of the organic compounds from food samples is an important application field of SDME technique, it earns nowadays great interest proved by the published papers in recent years. Overview of papers reporting the analysis of various samples, such as tea (Liu et al. 2012; Wu et al. 2015), wine (Garbi et al. 2010; Perreira Dos Anjos et al. 2015), vegetables (Amrvrazi and Tsiropoulos 2009; Kin et al. 2009; Wu et al. 2016) and fruits (Perreira Dos Anjos et al. 2014; Pano- Farias et al. 2017) are summarized in the Table 3. All of them were targeted on the determination of the pesticides. No special pre-treatment was reported for liquid samples such as wine or coconut water, mainly conventional sonication, acidification, pH adjustment and centrifugation were used (Garbi et al. 2010; Perreira Dos Anjos et al. 2014; Perreira Dos Anjos et al. 2015). On the other hand, more difficult pre-treatment was required for solid samples prior to the extraction step (Amrvrazi and Tsiropoulos 2009; Kin et al. 2009; Liu et al. 2012; Wu et al. 2015; Wu et al. 2016; Pano-Farias et al. 2017). DI was the most occurred mode of SDME (Amrvrazi and Tsiropoulos 2009; Garbi et al. 2010; Perreira Dos Anjos et al. 2014; Perreira Dos Anjos et al. 2015; Pano-Farias et al. 2017). For the extraction of 8 pesticides from cucumber and strawberries, the HS mode was selected and its application was compared to headspace solid phase microextraction (HS-SPME) and solid phase extraction (SPE) (Kin et al. 2009). More complex apparatus was employed by Wu et al. (2015, 2016) for the dynamic microwave assisted extraction (DMAE) and dynamic microwave assisted extraction in combination with continuous flow microextraction (DMAE-CFME) of pesticides from tea (Wu et al. 2015) and cabbage, cauliflower, red cabbage and cucumber (Wu et al. 2016). Furthermore, DSDME using 100 μL isooctane for isolation of OCPs was used to analyze samples of tea by Liu et al. (2012). Bereitgestellt von Slovenská poľnohospodárska knižnica | Heruntergeladen 28.02.20 08:24 UTC Nova Biotechnol Chim (2018) 17(1): 1-15 9 W u e t a l. (2 0 1 5 ) L iu e t a l. (2 0 1 2 ) Pe rr e ir a D o s A n jo s e t a l. (2 0 1 5 ) G a rb i e t a l. (2 0 1 0 ) W u e t a l. (2 0 1 6 ) Pe rr e ir a D o s A n jo s e t a l. (2 0 1 4 ) K in e t a l. (2 0 0 9 ) Bereitgestellt von Slovenská poľnohospodárska knižnica | Heruntergeladen 28.02.20 08:24 UTC Nova Biotechnol Chim (2018) 17(1): 1-15 10 A m rv ra z i a n d T si ro p o u lo s (2 0 0 9 ) P a n o -F a ri a s e t a l. ( 2 0 1 7 ) K u m a ri e t a l. (2 0 1 6 ) P e li t a n d Y e n g in (2 0 1 4 ) T a b le 2 . Bereitgestellt von Slovenská poľnohospodárska knižnica | Heruntergeladen 28.02.20 08:24 UTC Nova Biotechnol Chim (2018) 17(1): 1-15 11 For effective extraction of the pesticides from the vegetable samples and stable drop in the continuous flow by DMAE-CFME, a variety of water-immiscible extraction solvents (n-hexane, cyclohexane, dichloromethane, chloroform, carbon tetrachloride, ethyl acetate, and toluene) were investigated, resulting dichloromethane and ethyl acetate as inappropriate for this mode. The major loose of these solvents was presented in the continuous sample flow arrangement. Toluene exhibited the most promising results of extraction efficiency for the majority of studied compounds (Wu et al. 2016). The wide scale of solvents, such as dichloromethane, chloroform, carbon tetrachloride, and chlorobenzene were examined for the extraction of the pesticides from the tea. An improved extraction chamber using as a holder of the microdrop was employed. The microdrop provided a sufficient stability in the flowing system and the microdrop transfer to analytical instruments by microsyringe after extraction was without loose of the extraction solvent. Carbon tetrachloride exhibited the highest extraction efficiency for most of the analytes, also the carbon tetrachloride extract was proved to be cleaner in comparison to other solvents. Chlorobenzene was not stable enough in the flow (Wu et al. 2015). Analyzing tea samples, cyclohexane drop immersed to the tea sample solution was not reproducible with low stability. Toluene was problematic as it extracted tea pigments and further cleaning was necessary. Isooctane showing the highest viscosity among studied solvents was selected as the final solvent for real-samples extraction, in addition, it formed stable microdrop (Liu et al. 2012). Microsyringe of the volume 10 μL was used in all of the mentioned applications and the drop of the sample was in the range of 1 μL and 10 μL. Especially, the volume 100 μL was used for the DSDME mode to measure the volume of the sample. Presence of the steady vortex was important for the formation of the microdrop (Liu et al. 2012). To avoid carry over effect, the rinse of the microsyringe was used mostly by the extraction solvent or polar organic solvents (Amrvrazi and Tsiropoulos 2009; Kin et al. 2009; Garbi et al. 2010; Perreira Dos Anjos et al. 2015; Pano-Farias et al. 2017). The highest speed of the magnetic stirrer was used for the DSDME extraction, the stirring speed above 1,100 rpm causes instability and dissolution of the solvent microdrop (Liu et al. 2012). Low speed of magnetic stirrer was mostly used for the DI-SDME in the range of 180 – 700 rpm (Amrvrazi and Tsiropoulos 2009; Garbi et al. 2010; Perreira Dos Anjos et al. 2014; Perreira Dos Anjos et al. 2015; Pano-Farias et al. 2017) and microwave irradiation was employed for the DMAE (Wu et al. 2015) and DMAE-CFME (Wu et al. 2016). The influence of microwave energy (100 – 350 W) on the extraction was studied. Implementation of the microwave-assisted extraction (MAE) prior to SDME simplify the pre-treatment process by the acceleration of the disruption of the vegetable cells under the high temperature. The high temperature improves the diffusion and mass transfer during extraction, and the dissolving capacity of extraction solvent at the same time, as well as, the target compounds dissolves faster. However, when high microwave power is used, it may cause the analyte degradation (Wu et al. 2016). No salt addition was reported in the published papers for food samples analysis. Satisfactory recovery values in the acceptable limit (between 70 % and 120 %) were mostly reposted. Difficulties to extract some compounds were mostly explained by chemical properties of the analytes and by the interaction of the analytes with other compounds presented in the sample. Low values of recovery were reported for dursban (the pesticide) with recoveries in the interval of 13.2 % – 52.4 % (Perreira Dos Anjos et al. 2015), and for the pyrethroid insecticide λ-cyhalothrin (1.4 for grape and 1.5 for apple) (Amrvrazi and Tsiropoulos 2009). Permethrin I and permethrin II are compounds exhibiting higher affinity toward water phase, which is demonstrated by low partition coefficient, therefore, as a result, insufficient recovery rates were reported (39.4 % and 45.0 %, respectively). Thus, it is possible to notice a difficulty of the migration of these analytes from the aqueous sample to the organic solvent. Although a relatively high partition coefficient belonging to the pesticide endosulfane, a low recovery rate was reported (29.0 %), what was explained by the interaction with other matrix components comprising the coconut water. In general, extraction of the compounds with higher polarity such as OCPs, compared to other classes of compounds such as OPPs, is more difficult Bereitgestellt von Slovenská poľnohospodárska knižnica | Heruntergeladen 28.02.20 08:24 UTC Nova Biotechnol Chim (2018) 17(1): 1-15 12 hence they promote a higher affinity for water- based sample than for organic extractant (Perreira Dos Anjos et al. 2014). Biological samples Sample preparation step is essential for the isolation of analytes from complex biological matrices and has a great influence on their reliable and accurate determination. Suitable extraction technique is usually necessary because almost all biological samples are incompatible with the following chromatographic instrumentation, these samples exhibit complicated matrix for direct analysis, and components in the sample may interfere the signal acquisition. In addition, the analytes are present at concentration levels below the limit of detection of regular analytical methods (Ocaña-González et al. 2016). Biological samples contain proteins, salts, acids, bases, and various organic components with properties close to those of searched analytes. In the recent years, applications of the SDME were aimed mostly to extraction of the several organic compounds in the biological samples such as fish (Botrel et al. 2017), urine and blood of the rats (Agrawal et al. 2007), human urine (Gao et al. 2011; García-Vasquéz et al. 2015), saliva (Timofeeva et al. 2016) and human blood serum (Shrivas and Patel 2009) (Table 4). The solidification of the extraction solvent in microliter scale after dispersive liquid-liquid microextraction for the extraction of drugs (Ebrahimzadeh et al. 2011; Ahmadi-Jouibari et al. 2014; Tehrani et al. 2014), medicaments (Ebrahimzadeh et al. 2013; Jia et al. 2013; Suh et al. 2014) was also reported. A few papers were devoted to the extraction of pesticides, namely maneb (fungicide) and paraque (herbicide) in albine mice by Kumari et al. (2016) and chlorpyrifos and chlorpyrifos- oxon by Pelit and Yengin (2014). 1-Undecanol, 1-dodecanol, 2-dodecanol and n-hexadecane were investigated for the extraction of maneb and paraque from albine mice. Except n-hexadecane, all other solvents were found to recover comparably the same amount of targeted analytes. This maybe because all these solvents possess nearly similar affinity toward the targeted analytes. In this case, the solidification of the solvent drop was used, which offered manipulation with a microdrop without significant losing of extract. This approach was reported to be semi- automated and could be fully automated without varying from the original concept by the further modifications. Finally, 1-dodecanol was chosen to be used for further experiments, due to its peculiar characteristic of being easily solidified near room temperature. (Kumari et al. 2016). In the case of urine, this matrix offers several advantages as it is very easy to obtain and specimen volume is not a limitation. Biomonitoring of the chlorpyrifos exposure in human urine was reported and 2-dodecanol as extraction solvent was selected due to the absence of the solvent peak presented in the same time as the analytes in the chromatogram in comparison with the other studied solvents. The extraction was performed using increased temperature, and no addition of the salt was presented. In this case, it was difficult to collect the organic phase due to the formation of smaller droplets of the solvent in the solution after addition of 0.5 g NaCl to 10.0 mL of the solution (Pelit and Yengin 2014). Conclusions SDME showed its suitability for the extraction of various type of analytes, which belongs to the wide group of harmful organic compounds and environmental pollutants from numerous types of samples. Applications of SDME for real samples analysis were sorted into 3 groups according to sample nature and summarized into the form of overview tables. As it was shown by the largest group of applications, SDME has the major utilization in the area of environmental samples analysis, then followed by food samples analysis. Nowadays, there is an increasing frequency of publication devoted to the SDME for biological samples analysis. SDME offered the reach of high pre-concentration, clean final extract directly suitable for analysis, fast extraction and the simplicity of the process. It was shown, that the two most common SDME modes DI and HS have slightly different general fields of applicability, but they show their wide applicability for various matrices. DI-SDME was approved to be suitable mostly for the extraction of the nonpolar or moderately polar volatile and semivolatile analytes from relatively clean matrices, such as Bereitgestellt von Slovenská poľnohospodárska knižnica | Heruntergeladen 28.02.20 08:24 UTC Nova Biotechnol Chim (2018) 17(1): 1-15 13 water, using appropriate water immiscible solvents, due to the direct contact of the solvent with a water sample. Since volatile compounds are best pre- concentrated by headspace SDME. The selection of the working parameter was also established to be critical for the reaching satisfactory recoveries of the analytes and explained by numerous examples. In conclusion, miniaturized extraction methods such as SDME represent a new approach that is currently receiving a great deal of interest of researchers in the area of sample preparation methods for analytical purposes. Acknowledgements This work was supported by the Slovak Research and Development Agency under the contract No. SK-SRB- 2016-0006. SZ would like to thank for financial contribution from the STU Grant scheme for Support of Young Researchers. References Agrawal K, Wu FH (2007) Drop-to-drop solvent microextraction coupled with gas chromatography/mass spectrometry for rapid determination of trimeprazine in urine and blood of rats: application to pharmacokinetic studies. Rapid Commun. Mass Spectrom. 21: 3352-3356. Ahmadi F, Assadi Y, Hosseini SM, Rezaee M (2006) Determination of organophosphorus pesticides in water samples by single drop microextraction and gas chromatography-flame photometric detector. J. Chromatogr. A 1101: 307-312. Ahmadi-Jouibari T, Fattahi N, Shamsipur M (2014) Rapid extraction and determination of amphetamines in human urine samples using dispersive liquid-liquid microextraction and solidification of floating organic drop followed by high performance liquid chromatography. J. Pharm. Biomed. Anal. 94: 145-151. Alexovič M, Horstkotte B, Solich P, Sabo J (2016) Automation of static and dynamic non-dispersive liquid phase microextraction. Part 1: Approaches based on extractant drop-, plug-, film- and microflow- formation. Anal. Chim. Acta 906: 22-40. Amde M, Tan ZQ, Liu R, Liu JF (2015) Nanofluid of zinc oxide nanoparticles in ionic liquid for single drop liquid microextraction of fungicides in environmental waters prior to high performance liquid chromatographic analysis. J. Chromatogr. A 1395: 7-15. Amvrazi EG, Tsiropoulos NG (2009) Application of single- drop microextraction coupled with gas chromatography for the determination of multiclass pesticides in vegetables with nitrogen phosphorus and electron capture detection. J. Chromatogr. A 1216: 789-2797. Andraščíková M, Hrouzková S (2016) Fast preconcentration of pesticide residues in oilseeds by combination of QuEChERS with dispersive liquid–liquid microextraction followed by gas chromatography-mass spectrometry. Food Anal. Meth. 9: 2182-2193. Botrel BMC, Abreu DCP, Saczk AA, Bazana MJF, Coelho SM, El Rosa PV, De Lima RMF (2017) Residual determination of anesthetic menthol in fishes by SDME/GC-MS. Food Chem. 229: 674-679. Carlos EA, Alves RD, De Queiroz MEL, Neves A (2013) Simultaneous determination of the organochlorine and pyrethroid pesticides in drinking water by single drop microextraction and gas chromatography. J. Braz. Chem. Soc. 24: 1217-1227. De Souza Pinheiro A, De Andrade JB (2009) Development, validation and application of a SDME/GC-FID methodology for the multiresidue determination of organophosphate and pyrethroid pesticides in water. Talanta 79: 1354-1359. De Souza Pinheiro A, Da Rocha GO, De Andrade JB (2011) A SDME/GC–MS methodology for determination of organophosphate and pyrethroid pesticides in water. Microchem. J. 99: 303-308. Ebrahimzadeh H, Dehghani Z, Asgharinezhad AA, Shekari N, Molaei K (2013) Determination of haloperidol in biological samples with the aid of ultrasound‐assisted emulsification microextraction followed by HPLC‐DAD. J. Sep. Sci. 36: 1597-1603. Ebrahimzadeh H, Saharkhiz Z, Tavassoli M, Kamarei F, Asgharinezhad AA (2011) Ultrasound‐assisted emulsification microextraction based on solidification of floating organic droplet combined with HPLC‐UV for the analysis of antidepressant drugs in biological samples. J. Sep. Sci. 34: 1275-1282. Gao W, Chen G, Chen Y, Zhang X, Yin Y, Hu Z (2011) Application of single drop liquid–liquid–liquid microextraction for the determination of fluoroquinolones in human urine by capillary electrophoresis. J. Chromatogr. B 879: 291-295. Garbi A, Sakkas V, Fiamegos YC, Stalikas CD, Albanis T (2010) Sensitive determination of pesticides residues in wine samples with the aid of single-drop microextraction and response surface methodology. Talanta 82: 1286-1291. García‐Vázquez A, Borrull F, Calull M, Aguilar C (2016) Single‐drop microextraction combined in‐line with capillary electrophoresis for the determination of nonsteroidal anti‐inflammatory drugs in urine samples. Electrophoresis 37: 274-281. Jain A, Verma KK (2011) Recent advances in applications of single-drop microextraction: a review. Anal. Chim. Acta 706: 37-65. Jeannot MA, Cantwell FF (1996) Solvent microextraction into a single drop. Anal. Chem. 68: 2236-2240. Jeannot MA, Przyjazny A, Kokosa JM (2010) Single drop microextraction-development, applications and future trends. J. Chromatogr. A 1217: 2326-2336. Jia S, Li J, Park SR, Ryu Y, Park IH, Park JH, Lee J (2013) Combined application of dispersive liquid–liquid microextraction based on the solidification of floating Bereitgestellt von Slovenská poľnohospodárska knižnica | Heruntergeladen 28.02.20 08:24 UTC Nova Biotechnol Chim (2018) 17(1): 1-15 14 organic droplets and charged aerosol detection for the simple and sensitive quantification of macrolide antibiotics in human urine. J. Pharm. Biomed. Anal. 86: 204-213. Kaykhaii M, Saeed N, Chamsaz M (2005) Determination of aliphatic amines in water by gas chromatography using headspace solvent microextraction. Talanta 65: 223-228. Kin ChM, Tan GH (2009) Comparison of HS-SDME with SPME and SPE for the determination of eight organochlorine and organophosphorus pesticide residues in food matrices. J. Chromatogr. Sci. 47: 694-699. Kumari R, Jha RR, Singh MP, Patel DK (2016) Whirling agitated single drop microextraction technique for the simultaneous analysis of Paraquat and Maneb in tissue samples of treated mice. J. Sep. Sci. 39: 1725-1733. Lamboropoulou DA, Albanis TA (2007) Liquid-phase micro- extraction techniques in pesticide residue analysis. J. Biochem. Biophys. Methods 70: 195-228. Liu S, Dasgupta PK (1995) Liquid droplet. A renewable gas sampling interface. Anal. Chem. 7: 2042-2049. Liu S, Dasgupta PK (1996) Analytical chemistry in a drop. Solvent extraction in a microdrop. Anal. Chem. 68: 1817-1821. Liu D, Min S (2012) Rapid analysis of organochlorine and pyrethroid pesticides in tea samples by directly suspended droplet microextraction using a gas chromatography–electron capture detector. J. Chromatogr. A 1235: 166-173. Lopez-Blanco MC, Blanco-Cid S, Cancho-Grande B, Simal- Gandara J (2003) Application of single-drop microextraction and comparison with solid-phase microextraction and solid-phase extraction for the determination of α- and β-endosulfan in water samples by gas chromatography–electron-capture detection. J. Chromatogr. A 984: 245-252. Ocaña-González JA, Fernández-Torres R, Bello-López MÁ, Ramos-Payán M (2016) New developments in microextraction techniques in bioanalysis. A review. Anal. Chim. Acta 905: 8-23. Pano-Farias NS, Ceballos-Magana SG, Muniz-Valencia R, Jurado JM, Alcázar Á, Aguayo-Villarreal IA (2017) Direct immersion single drop micro-extraction method for multi-class pesticides analysis in mango using GC– MS. Food Chem. 237: 30-38. Pelit FO, Yengin Ç (2014) Application of solidified floating organic drop microextraction method for biomonitoring of chlorpyrifos and its oxon metabolite in urine samples. J. Chromatogr. B 949: 109-114. Pena-Pereira F, Lavilla I, Bwndicho C (2010) Liquid-phase microextraction techniques within the framework of green chemistry. Trends Anal. Chem. 29: 617-628. Perreira Dos Anjos J, De Andrade JB (2014) Determination of nineteen pesticides residues (organophosphates, organochlorine, pyrethroids, carbamate, thiocarbamate and strobilurin) in coconut water by SDME/GC–MS. Microchem. J. 112: 119-126. Perreira Dos Anjos J, De Andrade JB (2015) Simultaneous determination of pesticide multiresidues in white wine and rosé wine by SDME/GC-MS. Microchem. J. 120: 69-76. Przyjazny A, Kokosa JM (2002) Analytical characteristics of the determination of benzene, toluene, ethylbenzene and xylenes in water by headspace solvent microextraction. J. Chromatogr. A 977: 143-153. Ramos L (2012) Critical overview of selected contemporary sample preparation techniques. J. Chromatogr. A 1221: 84-98. Salemi A, Rasoolzadeh R, Nejad MM, Vosough M (2013) Ultrasonic assisted headspace single drop micro- extraction and gas chromatography with nitrogen- phosphorus detector for determination of organophosphorus pesticides in soil. Anal. Chim. Acta 769: 121-126. Santos LO, Dos Anjos JP, Ferreira SL, De Andrade JB (2017) Simultaneous determination of PAHS, nitro-PAHS and quinones in surface and groundwater samples using SDME/GC-MS. Microchem. J. 133: 431-440. Saraji M, Bakhshi M (2005) Determination of phenols in water samples by single-drop microextraction followed by in-syringe derivatization and gas chromatography–mass spectrometric detection. J. Chromatogr. A 1098: 30-36. Shrivas K, Patel DK (2009) Matrix-assisted laser desorption/ionization mass spectrometry for quantitative determination of β-blocker drugs in one-drop of human serum sample. J. Chromatogr. B 879: 35-40. Soares CES, Neves AA, Quieroz ME, Oliveira AF, Costa AI, Assis RC, Andrade CEO (2014) Single drop microextraction: A sensitive multiresidue method for determination of pesticides in water using GC/ECD. J. Braz. Chem. Soc. 25: 2016-2025. Soares CE, Neves AA, Queiroz ME, Oliveira AF, Costa AI, Assis RC, Andrade CEO (2015) Determination of pesticides in soil using a hyphenated extraction technique. J. Braz. Chem. Soc. 26: 1790-1797. Suh JH, Lee YY, Lee HJ, Kang M, Hur Y, Lee SN, Han SB (2014) Dispersive liquid–liquid microextraction based on solidification of floating organic droplets followed by high performance liquid chromatography for the determination of duloxetine in human plasma. J. Pharm. Biomed. Anal. 75: 214-219. Tehrani MS, Givianrad MH, Mahoor N (2014) Surfactant- assisted dispersive liquid–liquid microextraction followed by high-performance liquid chromatography for determination of amphetamine and methamphetamine in urine samples. Anal. Methods 4: 1357-1364. Tian F, Liu W, Fang H, An M, Duan S (2014) Determination of six organophosphorus pesticides in water by single- drop microextraction coupled with GC-NPD. Chromatographia 77: 487-492. Timofeeva I, Medinskaia K, Nikolaeva L, Kirsanov D, Bulatov A (2016) Stepwise injection potentiometric determination of caffeine in saliva using single-drop microextraction combined with solvent exchange. Talanta 150: 655-660. Yohannes A, Tolesa T, Merdassa Y, Megersa N (2016) Bereitgestellt von Slovenská poľnohospodárska knižnica | Heruntergeladen 28.02.20 08:24 UTC Nova Biotechnol Chim (2018) 17(1): 1-15 15 Single drop microextraction analytical technique for simultaneous separation and trace enrichment of atrazine and its major degradation products from environmental waters followed by liquid chromatographic determination. J. Anal. Bioanal. Tech. 7: 1-8. Wang X, Li Y, Cai F, Qing Q, Juan K, Chen B, Luan T (2017) Fully automatic single-drop microextraction with one-setp extraction and derivatization and its application for rapid analysis of hydroxylated polycyclic aromatic hydrocarbons in seawaters. Talanta 164: 727-734. Williams DBG, George MJ, Marjanovič L (2014) Rapid detection of atrazine and metolachlor in farm soils: gas chromatography–mass spectrometry-based analysis using the bubble-in-drop single drop microextraction enrichment method. J. Agri. Food Chem. 62: 7676-7681. Wu Y, Hu B, Hu W, Jiang Z, Li B (2007) A novel capillary microextraction on ordered mesoporous titania coating combined with electrothermal vaporization inductively coupled plasma mass spectrometry for the determination of V, Cr and Cu in environmental and biological samples. J. Mass. Spectrom. 42: 467-475. Wu L, Hu M, Li Z, Song Y, Zhang H, Yu A, Wang Z (2015) Dynamic microwave-assisted extraction online coupled with single drop microextraction of organophosphorus pesticides in tea samples. J. Chromatogr. A 1407: 42-51. Wu L, Hu M, Li Z, Song Y, Yu C, Zhang H, Wang Z (2016) Dynamic microwave-assisted extraction combined with continuous-flow microextraction for determination of pesticides in vegetables. Food Chem. 192: 596-602. Wu Y, Xia L, Chen R, Hu B (2008) Headspace single drop microextraction combined with HPLC for the determination of trace polycyclic aromatic hydrocarbons in environmental samples. Talanta 74: 470-477. Xie HY, Yan J, Jahan S, Zhong R, Fan LY, Xiao H, Cao CX (2014) A new strategy for highly efficient single-drop microextraction with a liquid–gas compound pendant drop. Analyst 139: 2545-2550. Bereitgestellt von Slovenská poľnohospodárska knižnica | Heruntergeladen 28.02.20 08:24 UTC