S4

S Afr Fam Pract
ISSN 2078-6190    EISSN 2078-6204 

© 2019 The Author(s)

NW REFRESHER COURSE

Summary

PMs and ICDs are cardiac implantable electronic devices 
(CIEDs) that are becoming increasingly sophisticated and the 
perioperative management of these devices is changing along 
with this development. Traditionally, PM functions have been 
changed to asynchronous modes during surgery because of 
the fear of electromagnetic interference (EMI) from diathermy 
causing oversensing and subsequent loss of pacing. ICDs have 
been switched to off mode to prevent inadvertent shocks during 
EMI. This may lead to patient harm, due to R-on-T phenomenon 
in PM set in asynchronous mode and undiagnosed perioperative 
v-tach or v-fib in patients who have ICDs in off mode. PM-on and 
ICD-on strategies are becoming more acceptable, depending 
on the site of surgery. Intraoperative magnet use is currently 
underutilised and may have advantages to changing PM and 
ICD settings in patients who may otherwise have had the CIED 
functions switched off. Reversal of functions to preoperative 
settings may be achieved in the operating theatre without the 
need of a PM technologist. 

Background

Approximately 7500 PMs and ICDs are implanted every year into 
patients in South Africa. Discovery Medical Aid submissions for 
PMs and ICDs number 7200 per year, but this figure includes 
new devices, temporary pacemaker insertion, generator 
replacement for end of service (EOS) or end of life (EOL) and lead 
changes or repositioning (see Figure 1).1,2 With patient longevity 
rates increasing, it is becoming more likely that anaesthetists 
will encounter patients with PMs and ICDs, especially in the 
elderly. The medical technologists who specialise in CIEDs 
provide an important service in interrogating these devices and 
anaesthetists need to work closely with them in providing the 
optimum perioperative care when patients with devices present 
for surgery.  All patients presenting for surgery need a recent 
interrogation of their device and the information presented 
to the surgical team. The anaesthetist should make decisions 
regarding pacemaker management in conjunction with the 
technologist, cardiologist and surgeon (see Figure 4). 

Modern PMs and ICDs

PM technology was developed 60 years ago and recent 
advances, especially over the last 10 years have taken a relatively 

crude device that initially provided an asynchronous pacing 
beat to the right ventricle as a life saving procedure for patients 
with complete heart block to the sophisticated modern device 
that paces on demand, reacts to exercise by increasing the 
heart rate to a pre-set level and in the operating environment 
can distinguish between a sensed beat and EMI. Many of the 
guidelines for the management of PMs and ICDs were published 
at the beginning of the current decade and could do with a 
complete revision, taking into account the developments of 
these devices over the last 10 years.

Perioperative considerations

PMs The anaesthetist has to weigh up the pros and cons of 
either leaving a PM with its usual settings (PM-on) or asking 
the technologist to change the settings to asynchronous mode 
(either DOO or VOO). The problem that may be experienced with 
a PM left in DDD mode is oversensing due to electromagnetic 
interference (EMI) when the PM senses the EMI as electrical 
activity of the heart and inhibits the generation of a paced 
beat. This may result in periods of asystole while diathermy is 
being used. Modern PMs have algorithms that can distinguish 
between EMI and normal electrical activity in the heart and 
may ignore diathermy induced EMI completely or change 
the PM setting temporarily to DOO. The problem of a PM 
changed to asynchronous mode is that an R-on-T phenomenon 
may occur, which may result in ventricular fibrillation if an 
asynchronous beat is delivered during the refractory period 
of the cardiac cycle. The benefits of a PM-on protocol are that 
the chances of an undiagnosed R-on-T phenomenon occurring 
perioperatively are reduced and in addition, that the PM settings 
do not have to be reset by a technologist postoperatively.  If 
the anaesthetist is planning to follow a PM-on protocol, the 
preoperative interrogation should include the PM response 
to magnet application as well as the ability to apply a magnet 
intraoperatively to manage oversensing if it occurs due to EMI 
(see Figure 7).

Many pacemakers are implanted today for disease of the 
sinu-atrial node (Sick sinus syndrome - SSS) where the patient 
experiences syncope due to bradycardia associated with SSS. 
These patients have normal atrioventricular conduction and the 
PM is set on AAI. Changing the PM to DOO or VOO may result 
in asynchronous ventricular contraction because the ventricular 

South African Family Practice 2019; 61(2):S4-S8
 
Open Access article distributed under the terms of the 
Creative Commons License [CC BY-NC-ND 4.0] 
http://creativecommons.org/licenses/by-nc-nd/4.0

Perioperative Management of Pacemakers (PM) and Implantable Cardioverter 
Defibrillators (ICD) in South Africa
Keene A 

Private Practice, Johannesburg 
Corresponding author, email: tandpkeene@mweb.co.za



Perioperative Management of Pacemakers (PM) and Implantable Cardioverter Defibrillators (ICD) in South Africa 5

S5

lead is usually placed in the right ventricle. The left ventricular 
conduction and contraction is delayed and this may result in 
reduced left ventricular efficiency. These patients would benefit 
from a PM-on protocol.

ICDs are potentially problematic in that an ICD-on protocol 
may result in unwanted and repeated shocks due to EMI being 
misinterpreted by the device as ventricular fibrillation. This can 
lead to myocardial damage and depletion of battery life. It has 
been shown that ICDs do not respond to EMI if the operation site 
is below the iliac crest and that the diathermy dispersive pad is 
placed at a site to lead current away from the ICD.3 If an ICD-on 
protocol is to be used, the anaesthetist should be aware of the 
response to application of a magnet to the specific device (see 
Figure 8).

An ICD-off protocol may result in undiagnosed v-tach or 
v-fib in the perioperative period and case reports have been 
published where patients have died at home after the 
device was not checked and turned on postoperatively. 
International literature is clear that the responsibility for 
resetting the ICD belongs to the anaesthetist and not 
to the technologist.4 An ICD-off protocol may result in 
unnecessary external cardioversion for v-tach because 
anti-tachy pacing (ATP) is disabled. ATP reduces the 
need for shocks in patients who develop v-tach as many 
tachyarrhythmias can be terminated by the rapid pacing 
before the device has to deliver a shock (see Figure 3).

The standardised pacemaker codes are depicted in 
Figure 2.5 Multisite ventricular pacing refers to cardiac 
resynchronisation therapy (CRT) for the left and right 
ventricles. It is utilised in patients with left bundle 
branch block to resynchronise left and right ventricular 
contraction to occur simultaneously. This may improve 
cardiac output by up to 15% in patients with heart failure 
due to reduced ejection fraction. Multisite atrial pacing is 
used experimentally to treat atrial fibrillation.6

Implantable Cardioverter Defibrillator (ICD) and 
Anti-Tachycardia Pacing (ATP)

Patients with recurrent ventricular tachycardia and/or ventricular 

fibrillation or those at risk for developing these arrhythmias may 

have an implantable cardioverter defibrillator (ICD) placed in the 

left subclavian region. The ICD senses the R-R interval and if the 

interval reduces to a predetermined level, the device algorithm 

reads this as ventricular tachycardia and can deliver a repetitive 

sequence of eight rapid paced beats to try to break the re-entry 

condition of v-tach. If this fails to cardiovert the v-tach, a high 

voltage shock is delivered (see Figure 3).  The shock is delivered 

from the coils to the generator in a triangulated vector to 

incorporate the left ventricle. (The high voltage coils around the 

pacemaker leads act as the cathode and the pulse generator 

acts as the anode). ICDs recognise supraventricular tachycardias 

(SVTs, AF and sinus tachycardia) via atrial sensing, but cannot 

cardiovert them. This is designed to prevent unnecessary shocks. 

ICDs may deliver anti-bradycardia therapy if required to do 

so.7 A magnet applied to an ICD will generally disable the anti-

tachycardia therapy while it is in situ, but will have no effect on 

anti-bradycardia therapy or rate responsiveness. 

Figure 1: Implantation rate estimated to be 138.25/million in 2016 (total 
population 54146735, total number 7485) US rates are double Germany

Figure 2: Pacemaker Codes

Figure 3: De-Identified printout from a patient’s ICD showing an episode of V-Tach 
with attempted anti-tachy pacing (ATP) followed by successful 34.5 J defibrillation. 
Image supplied by Medtronic SA.



S Afr Fam Pract 2019;61(2)6

S6

Preoperative pacemaker interrogation is considered to be the 

standard of care and should be scheduled during the week prior 

to surgery. The PM technologist should provide the information 

as shown in Figure 4 and in consultation with the technologist 

and cardiologist and taking into account the planned surgery, 

the anaesthetist should decide on the perioperative PM 

management (see Figure 7).

If a PM-on strategy is to be followed, it is vitally important that 

the effects of magnet application and removal to the specific 

PM are known. It may be advantageous to reset the base rate 

of the pacemaker, even though a PM-on protocol is followed. 

For example, a base rate of 60 may be increased to 70 or 80 if 

haemodynamic challenges are expected, such as blood loss or 

neuraxial anaesthesia.

If the PM settings are to be changed, this should be done on 

the day of surgery and preferably reversed to the preoperative 

settings as soon as possible after the ESU is no longer required. 

If an ICD-on strategy is to be followed, it is important to note that 

placing a magnet over the ICD will not change the underlying 

PM function and this has to be changed independently of the 

anti-tachyarrhythmia function. This should be considered in PM 

dependent patients if the surgery is close to the ICD and/or leads.  

CRT devices need to be checked within 3 months because the 
coronary sinus leads have poorer contact compared to the right 
ventricular leads and higher thresholds are accepted. This may 
result in faster battery drain and thus reduce longevity.

Response to magnet application

The response to magnet application in Medtronic PM and 
ICDs is shown in Figure 5. It is important to note that there is 
no uniformity across the industry, and each manufacturer has 
a different set of responses to magnet application. This makes 
it imperative that the anaesthetist is aware of the specific 
response of the patient’s device before applying a magnet. An 
indication of the enormity of this issue, is that in the USA, there 
are 1440 different types of device models across the different 
manufacturers.7,8 Modern PM and ICDs respond in a determined 
manner to magnet application, and many of the problems 
associated with application have been dealt with. Examples 
include resetting device programming under the influence 
of a magnet and concurrent EMI, and switching off anti-tachy 
functions in ICDs which do not return once the magnet is 
removed. These problems do not occur in modern devices. 

During surgery, magnet application needs to be carefully 
monitored. This is easily done in the patient with a PM, 
because magnet application will result in a fixed rate change, 
which is specific to the manufacturer. For example, Medtronic 
pacemakers change from the patient’s usual settings to a rate of 
85 beats per minute while the magnet is applied. This rate is 65 
if the battery life is shortened (EOL) and the generator needs to 
be replaced (RRT). Therefore, as long as the heart rate is 85 bpm, 
the anaesthetist can be reassured that the magnet is correctly 
applied.

It is less clear in the case of ICDs. The Medtronic ICDs emit a tone 
for 10 seconds when the magnet is applied, but after this initial 
signal, there is no indicator to the anaesthetist whether or not 
the magnet is still in place and exerting its effect on the ICD. The 
magnet has no effect on the PM function of the ICD, so heart rate 
changes cannot be used as an indicator. It would make sense for 
the manufacturers to provide the anaesthetist with an ongoing 
signal to determine correct placement of a magnet over an ICD. 
Modern devices have Bluetooth functionality and an elegant 

Patients with recurrent ventricular tachycardia and/or ventricular fibrillation or those at risk for 
developing these arrhythmias may have an implantable cardioverter defibrillator (ICD) placed in 
the left subclavian region. The ICD senses the R-R interval and if the interval reduces to a 
predetermined level, the device algorithm reads this as ventricular tachycardia and can deliver a 
repetitive sequence of eight rapid paced beats to try to break the re-entry condition of v-tach. If 
this fails to cardiovert the v-tach, a high voltage shock is delivered (see Figure 3).  The shock is 
delivered from the coils to the generator in a triangulated vector to incorporate the left ventricle. 
(The high voltage coils around the pacemaker leads act as the cathode and the pulse generator 
acts as the anode). ICDs recognise supraventricular tachycardias (SVTs, AF and sinus 
tachycardia) via atrial sensing, but cannot cardiovert them. This is designed to prevent 
unnecessary shocks. ICDs may deliver anti-bradycardia therapy if required to do so.7 A magnet 
applied to an ICD will generally disable the anti-tachycardia therapy while it is in situ, but will 
have no effect on anti-bradycardia therapy or rate responsiveness.  
 

 
 
Figure 4. Preoperative Pacemaker and ICD Interrogation 
 
Preoperative pacemaker interrogation is considered to be the standard of care and should be 
scheduled during the week prior to surgery. The PM technologist should provide the information 
as shown in Figure 4 and in consultation with the technologist and cardiologist and taking into 
account the planned surgery, the anaesthetist should decide on the perioperative PM 
management (see Figure 7). 
If a PM-on strategy is to be followed, it is vitally important that the effects of magnet application 
and removal to the specific PM are known. It may be advantageous to reset the base rate of the 
pacemaker, even though a PM-on protocol is followed. For example, a base rate of 60 may be 
increased to 70 or 80 if haemodynamic challenges are expected, such as blood loss or neuraxial 
anaesthesia. 
If the PM settings are to be changed, this should be done on the day of surgery and preferably 
reversed to the preoperative settings as soon as possible after the ESU is no longer required.  
If an ICD-on strategy is to be followed, it is important to note that placing a magnet over the ICD 
will not change the underlying PM function and this has to be changed independently of the anti-

Figure 5: Response to magnet application in Medtronic PM and ICDs

way off dealing with this problem would be to have a smartphone application that can read the 
wireless signals from the device. 
 

 
 
Figure 6: Operating Theatre Requirements 
 
PM-on or CRT-P-on protocol suggests that the PM settings are not changed during the standard 
preoperative technologist interrogation of the device. PM-on protocol will avoid inadvertent 
development of R-on-T phenomenon and implies that magnet application is possible if the 
situation of oversensing caused by EMI develops. In addition to this, the anaesthetist is able to 
remove the magnet when EMI is no longer being used. It takes away the need for postoperative 
high dependency unit care and a repeat call out by the technologist to reset the PM. The 
successful PM-on protocol requires the anaesthetist to follow the algorithm in Figure 7. A magnet 
can disable the rate responsiveness function in a PM and this is useful to prevent unwanted 
increases in heart rate, for example when the sternal saw is used in open-heart procedures, as 
the vibration of the saw may be misinterpreted as patient movement (see Figure 7). 
ICD or CRT-D-on protocol suggests that the anti-tachycardia therapy of the ICD is not switched 
off perioperatively and that the device is allowed to sense tachyarrhythmias and deliver anti-
tachycardia pacing (ATP) and shock if necessary. This provides protection to the patient in the 
event of V-Tach and/or V-fib, which may occur at anytime perioperatively. In order to prevent 
unnecessary ATP or shocks due to EMI being incorrectly read by the generator, a magnet should 
be available to place over the ICD to change the setting to ATP and Shock off. It is important to 
note that the anaesthetist should know exactly what happens to the specific ICD under the 
influence of magnet application and removal, that the patient position is such that the magnet can 
be properly secured during surgery and that the tone emitted by the generator is recognised. 
There are certain operating sites where it is safe to leave the ICD on and generally these are more 
that 15 cm  (6 inches) from the generator and leads and below the umbilicus or ileac crest, as 
long as the ESU dispersal pad is sited away from the surgical site so that EMI is not directed 
towards the ICD.3 For example, hip surgery is safe to proceed without switching off the anti-tachy 
therapy, as long as the dispersal pad is placed on the ipsilateral thigh.  
The anaesthetist should consider switching off the anti-tachy therapy of the ICD if:  
 

Figure 6: Operating Theatre Requirements

Figure 4. Preoperative Pacemaker and ICD Interrogation



Perioperative Management of Pacemakers (PM) and Implantable Cardioverter Defibrillators (ICD) in South Africa 7

S7

way of dealing with this problem would be to have a smartphone 
application that can read the wireless signals from the device.

PM-on or CRT-P-on protocol suggests that the PM settings are 
not changed during the standard preoperative technologist 
interrogation of the device. PM-on protocol will avoid 
inadvertent development of R-on-T phenomenon and implies 
that magnet application is possible if the situation of oversensing 
caused by EMI develops. In addition to this, the anaesthetist is 
able to remove the magnet when EMI is no longer being used. 
It takes away the need for postoperative high dependency unit 
care and a repeat call out by the technologist to reset the PM. 
The successful PM-on protocol requires the anaesthetist to 
follow the algorithm in Figure 7. A magnet can disable the rate 
responsiveness function in a PM and this is useful to prevent 
unwanted increases in heart rate, for example when the sternal 
saw is used in open-heart procedures, as the vibration of the saw 
may be misinterpreted as patient movement (see Figure 7).

ICD or CRT-D-on protocol suggests that the anti-tachycardia 
therapy of the ICD is not switched off 
perioperatively and that the device is allowed 
to sense tachyarrhythmias and deliver anti-
tachycardia pacing (ATP) and shock if necessary. 
This provides protection to the patient in the event 
of V-Tach and/or V-fib, which may occur at anytime 
perioperatively. In order to prevent unnecessary 
ATP or shocks due to EMI being incorrectly read 
by the generator, a magnet should be available to 
place over the ICD to change the setting to ATP 
and Shock off. It is important to note that the 
anaesthetist should know exactly what happens 
to the specific ICD under the influence of magnet 
application and removal, that the patient position 
is such that the magnet can be properly secured 
during surgery and that the tone emitted by 
the generator is recognised. There are certain 
operating sites where it is safe to leave the ICD 
on and generally these are more that 15 cm  (6 
inches) from the generator and leads and below 
the umbilicus or ileac crest, as long as the ESU 
dispersal pad is sited away from the surgical site 
so that EMI is not directed towards the ICD.3 For 
example, hip surgery is safe to proceed without 
switching off the anti-tachy therapy, as long as 
the dispersal pad is placed on the ipsilateral thigh. 

The anaesthetist should consider switching off 
the anti-tachy therapy of the ICD if: 

1. The operation site is within 15 cm of the 
generator or leads, especially if long bursts of 
unipolar diathermy are to be used. The argon 
beam ESU cannot be used in short bursts and 
may cause long periods of EMI.

2. A magnet cannot be reliably secured over the 
generator, such as in the prone position.

3. Certain operations such as hand surgery and 
ophthalmic operations where inadvertent 
shocks may lead to patient or operator harm. 

This would include operations at these sites performed under 

local anaesthesia. Thoracic operations would require the ICD-

off because left chest procedures would render the anti-tachy 

functions ineffective because of poor tissue shock transfer due 

to high impedance when the chest is open. Also, in right chest 

procedures, it would be difficult to ensure proper application 

and security of the magnet if it was required.9

If the patient is pacemaker dependent, the anaesthetist may 

consider asking the PM tech to change the PM settings to 

DOO as in the PM-on protocol. It is also advisable to turn off 

rate responsiveness, as this function is not changed by magnet 

application in ICDs.

It is important to note that if the anti-tachy function has been 

turned off (ICD-off ), the settings must be restored as soon as 

possible after the procedure and the patient has to be observed 

in a high dependency unit until this has been achieved.

1. The operation site is within 15 cm of the generator or leads, especially if long bursts of 
unipolar diathermy are to be used. The argon beam ESU cannot be used in short bursts 
and may cause long periods of EMI. 

2. A magnet cannot be reliably secured over the generator, such as in the prone position. 
3. Certain operations such as hand surgery and ophthalmic operations where inadvertent 

shocks may lead to patient or operator harm. This would include operations at these 
sites performed under local anaesthesia. Thoracic operations would require the ICD-off 
because left chest procedures would render the anti-tachy functions ineffective because 
of poor tissue shock transfer due to high impedance when the chest is open. Also, in right 
chest procedures, it would be difficult to ensure proper application and security of the 
magnet if it was required.9 

If the patient is pacemaker dependent, the anaesthetist may consider asking the PM tech to 
change the PM settings to DOO as in the PM-on protocol. It is also advisable to turn off rate 
responsiveness, as this function is not changed by magnet application in ICDs. 
It is important to note that if the anti-tachy function has been turned off (ICD-off), the settings 
must be restored as soon as possible after the procedure and the patient has to be observed in a 
high dependency unit until this has been achieved. 
 

 
 
Figure 7: PM-on algorithm 
 
 
 

 
Figure 8: ICD-on algorithm 
 
References 
 

1. Personal communication Darren Sweiden, Discovery.  
2. Bonny A, Ngantcha M, Jeilan M, Okello E, Kaviraj B, Talle MA, et al. Statistics on the use of 

cardiac electronic devices and interventional electrophysiological procedures in Africa 
from 2011 to 2016: report of the Pan African Society of Cardiology (PASCAR) Cardiac 
Arrhythmias and Pacing Task Forces. Europace. 2017;00:1-14. 

3. Gifford J, Larimer K, Thomas C, May P. ICD-ON Registry for Perioperative Management of 
CIEDs: Most Require No Change. PACE. 2017;40:128-34. 

4. Healey JS, Merchant R, Simpson C, Tang T, Beardsall M, Tung S, et al. Society Position 
Statement: Canadian Cardiovascular Society/ Canadian Anesthesiologists’ Society/ 
Canadian Heart Rhythm Society joint position statement on the perioperative 
management of patients with implanted pacemakers, defibrillators and 
neurostimulating devices. Can J Anesth. 2012 Apr;59(4):394-407 

5. Benson R. Pacemaker Nomenclature. St Louis Dept Anesthesia, 2005-2006. Available 
from: http://anesthesia.slu.edu/pdf/pacemaker.pdf 

6. Diprose P, Piece JM. Anaesthesia or patients with pacemakers and similar devices. BJA 
CEPD Reviews. 2001 Dec 1;1(6):166-70. 

7. Jacob S, Panaich SS, Maheshwari R, Haddad JW, Padanilam BJ, John SK. Clinical 
applications of magnets on cardiac rhythm management devices, Europace. 
2011:13(9):1222-30. 

8. Schulman PM, Rozner MA. Use caution when applying magnets to pacemakers or 
defibrillators for surgery. Anesth Analg. 2013;117:442-7. 

9. Crossley GH, Poole JE, Rozner MA, Asirvatham SJ, Cheng A, Chung MK, et al. The Heart 
Rhythm Society (HRS)/American Society of Anesthesiologists (ASA); Expert Consensus 
Statement on the perioperative management o patients with implantable defibrillators, 
pacemakers and arrhythmia monitors. Heart Rhythm. 2011 Jul;8(7):1114-54. 
 

Glossary of terms, abbreviations and acronyms 

Figure 7: PM-on algorithm

Figure 8: ICD-on algorithm



S Afr Fam Pract 2019;61(2)8

S8

References
1. Personal communication Darren Sweiden, Discovery. 

2. Bonny A,  Ngantcha M,  Jeilan M,  Okello E,  Kaviraj B, Talle MA, et al. Statistics on 
the use of cardiac electronic devices and interventional electrophysiological 
procedures in Africa from 2011 to 2016: report of the Pan African Society of 
Cardiology (PASCAR) Cardiac Arrhythmias and Pacing Task Forces. Europace. 
2017;00:1-14.

3. Gifford J,  Larimer K,  Thomas C,  May P. ICD-ON Registry for Perioperative 
Management of CIEDs: Most Require No Change. PACE. 2017;40:128-34.

4. Healey JS,  Merchant R,  Simpson C,  Tang T,  Beardsall M,  Tung S, et 
al. Society Position Statement: Canadian Cardiovascular Society/ Canadian 
Anesthesiologists’ Society/ Canadian Heart Rhythm Society joint position 
statement on the perioperative management of patients with implanted 
pacemakers, defibrillators and neurostimulating devices. Can J Anesth. 2012 
Apr;59(4):394-407

5. Benson R. Pacemaker Nomenclature. St Louis Dept Anesthesia, 2005-2006. 
Available from: http://anesthesia.slu.edu/pdf/pacemaker.pdf

6. Diprose P, Piece JM. Anaesthesia or patients with pacemakers and similar 
devices. BJA CEPD Reviews. 2001 Dec 1;1(6):166-70.

7. Jacob S,  Panaich SS,  Maheshwari R,  Haddad JW,  Padanilam BJ, John SK. Clinical 
applications of magnets on cardiac rhythm management devices, Europace. 
2011:13(9):1222-30.

8. Schulman PM, Rozner MA. Use caution when applying magnets to pacemakers 
or defibrillators for surgery. Anesth Analg. 2013;117:442-7.

9. Crossley GH,  Poole JE,  Rozner MA,  Asirvatham SJ,  Cheng A,  Chung MK, et al. 
The Heart Rhythm Society (HRS)/American Society of Anesthesiologists (ASA); 
Expert Consensus Statement on the perioperative management o patients with 
implantable defibrillators, pacemakers and arrhythmia monitors. Heart Rhythm. 
2011 Jul;8(7):1114-54.

Glossary of terms, abbreviations and acronyms

ATP – Anti Tachycardia Pacing

BOL –Beginning of Life

CIED – Cardiac Implantable Electronic Device

CRMD – Cardiac Rhythm Management Device

CRT – Cardiac Resynchronisation Therapy

EOL – End of Life

EOS – End of Service (same as RRT)

ESU – Electrosurgical Unit

ICD – Implantable Cardioverter Defibrillator 

IPD – Implantable Pulse Generator

PM – Pacemaker

RRT – Recommended Replacement Time