Upsala J Med Sci 82: 27-30, 1977 

On the Reproducibility of Exercise Tests in 
Patients with Atrial Fibrillation 

HANS &ERG, GUNNAR STROM and WAR WERNER 

From the Departments of Internal Medicine and Clinical Physiology, 
University Hospital, Uppsala, Sweden 

ABSTRACT 
Fifteen patients with atrial fibrillation, mostly due to 
advanced valvular heart disease, were examined by a 
graded work test and an orthostatic test on two occasions, 
the reason being, to study the reproducibility of the test in 
such patients. In 6 of the 7 patients with the highest working 
capacity the agreement between the two tests was good 
regarding heart rates at maximum work load. In the 8 
patients with a low working capacity the results of the two 
tests differed widely. In the orthostatic test the difference in 
increase of the heart rate after standing was less than 5 
beatslmin in 3 patients in one of the tests, and in 2 patients 
in both tests. The exercise test in patients with atrial fibrilla- 
tion must be evaluated with caution and has an acceptable 
reproducibility only in patients with fairly good functional 
capacity. 

INTRODUCTION 

We have previously reported on exercise tests in 
patients with atrial fibrillation (2, 3). 

The average increase in ventricular rates at 
gradually increased exercise loads was often found 
to be close to a linear repationship. In particular, 
this was the case in patients with a relatively well- 
maintained functional capacity of the heart. How- 
ever, in patients with more severely impaired heart 
function there was a tendency of a slightly more 
pronounced increase in the heart rate from resting 
value to especially that of the first load. In the 
literature different results have been presented on 
this problem (4, 7, 12). 

In another study we have investigated the influ- 
ence of different dose rates of digitalis on the heart 
rate and on the working capacity in patients with 
atrial fibrillation during exercise tests (3). By in- 
creasing the digitalis dosage the heart rate at the 
same load decreased. 

The aim of the present study was to test the 
reproducibility of orthostatic and exercise tests in 

patients with atrial fibrillation. In such an investiga- 
tion it is of critical importance that the cardiac state 
be stable and unchanged at, and in the interval 
between the two exercise tests. 

MATERIAL AND METHODS 
In 15 patients with atrial fibrillation exercise tests were 
done twice in each patient. The composition of the patient 
group is shown in Table I. The group is dominated by 
patients with valvular heart disease, often referred to this 
hospital for consideration for cardiac surgery. The exer- 
cise test was a part of this evaluation. Most patients had 
fairly advanced valvular disease. The repeated test was 
only performed with the patient’s consent since this test 
was made mainly for scientific reasons. 

This second exercise test was performed only in those 
patients fulfilling certain criteria to ensure a clinically sta- 
ble condition of the patient between the time of the two 
tests. These criteria were: 

(A) No change had occurred in clinical findings o r  in 
the patient’s history of his capacity between the two tests. 
The patient was not in heart failure. 

(B) The time interval between the two tests had not 
exceeded 3 months and i t  had frequently been much less. 
Seven of the cases had the tests less than 2 weeks apart. 

(C) The resting ECGs were identical on both occasions. 
In 3 patients with more than 1 month between the tests 
chest X-rays were performed before the two exercise 
tests and were found to be identical. 

(D) The patient had the same drugs in identical dosages 
on the two occasions. Serum potassium and other electro- 
lytes were within normal limits at both tests. 

(E) N o  other disease had occurred during the interval 
between the two tests. 

The orthostatic test preceding the exercise test was 
performed and described in more detail by Sandberg (9). 
According to the recurrent nomenclature the test was 
considered ‘negative’ when the heart rate increased less 
than 20 beatslmin and ‘positive’ if between 20 and 29. 

The exercise test was performed according to the 
graded ‘steady state’ principle (10, 13). The exercise test 
was performed using a bicycle ergometer (5, 6) with the 
method and calculations having previously been described 
(2). The heart rate was determined a t  rest, after 8 min 

Uosciki J M i 4  Sci 82 



28 H .  Aberg et al. 

Table I. Composition of the case material 

Diagnosis 

No. of patients Age Heart volume cc/m2 BSA 

Male Female Total Mean Range Mean Range 

Mitral stenosis 4 3 7 42 28-57 745 510-1 060 

(650) disease 1 1 2 (49) - - 

valve disease 2 3 5 48 39-62 730 590-1 010 
Unknown 1 1 (36) (550) - 

Combined mitral 

Mitral +aortic 

- 

Total 8 7 15 44.8 28-62 714 510-1 060 

standing, and after 2, 4 and 6 min a t  each load as the 
average of 25 consecutive heart intervals. The test was 
supervised by a physician at all times. 

The results of the exercise test was expressed in differ- 
ent ways: as the work load at a heart rate of ll0/min 
(W,,,), as the highest work load which according to clini- 
cal judgment of the patient’s symptoms and signs was 
permissible and possible to perform for 6 min (W,,,), and 
as the ability to reach ‘steady state’ as defined below. W,,, 
was calculated by slight interpolation or extrapolation, 
assuming a linear relationship between heart rate and 
work load within the small interval in question. In patients 
with atrial fibrillation and very low working capacity this 
is less precise (2). A ‘steady state’ was judged to be 
obtained when the difference between the heart rates a t  2, 
4 and 6 min on the same load was 10 beats/min or less. 

RESULTS AND COMMENTS 

The resting heart rates were fairly constant on the 
two occasions with few exceptions. In 13 of IS pa- 
tients the difference in resting heart rates were 10 
beats/min or less and in 6 of 15 only 5 beatslmin or 
less. The resting values (the lowest of the 2) were 
below 70 beats/min in 9 patients and above 90 (the 
highest of the 2) in 3 patients. These rates were 
rather high, considering that all patients except one 
was on maintenance dose digitalis. 

With regard to the orthostatic test there was a 
difference between the two tests in 3 patients (from 
‘positive’ t o  ‘negative’ or vice versa). Only in one 
was the difference greater. This patient had on the 
first occasion an increase of the heart rate of 26 
beatslmin after 8 min standing (‘positive’ reaction) 
and on the second test an increase of 6 beatslmin 
(‘negative’ reaction). The other 2 had 18 and 16 on 
the first and 23 and 25 on the second test, respec- 
tively. In S cases the difference was 5 beatslmin or 
less between the two tests of orthostatic reaction. 

The working capacity, measured as Wllo, was 
generally low, with a few exceptions (Table II). The 
mean W,,, was 168 (range 18439) and 191 (range 

Upsulu J Med Sci 82 

23497) kpm/min at the two exercise tests. This was 
expected, as many patients had fairly advanced 
heart disease. In 8 patients with a W,,, less than 100 
kpm/min in one or both of the exercise tests there 
were 4 patients whose difference between the two 
tests did not exceed 1/3 of the lowest value. In one 
the difference was about 40%. In the remaining 3 in 
this patient group with low working capacity the 
difference was as much as double or more the low 
W,,, in each pair. In the group with a working 
capacity W,,, above 110 kpmlmin, consisting of 7 
patients, the agreement between the two exercise 
tests was slightly better. In 3 patients the results 

Table 11. Results of exercise tests 
A is the first exercise test and B the second one. Defini- 
tions are given in Methods 

Patients 

w,,, W,,, 
(kpm/min) (kpmlmin) 

A B A B 

1 .  A . M .  
2. R . B .  
3 .  V . R .  
4 .  K . B .  
5 .  A . K .  
6. V. P. 
7. M. S. 
8. N .  E .  
9. G . H .  

10. E. 0. 
1 1 .  S. K. 
12. E . W .  
13. E. 0. 
14. G. S .  
15. S .  H .  
Mean 
S.D. 

18 23 
50 36 

177 169 
88 1 I6 

180 82 
23 30 

308 391 
167 86 
206 248 
75 260 

3 I7 497 
I92 139 
37 47 

239 376 
439 367 
168 191 
123 155 

150 
1 50 
200 
200 
200 
250 
250 
300 
400 
400 
400“ 
450 
500 
500 
700 
337 
158 

1 50 
1 50 
200 
200 
200 
250 
250 
300 
400 
400 
600 
500 
500 
500 
700 
3.53 
175 

a The patient showed aberrant ventricular ECG- 
complexes. The first test was then interrupted. In spite of 
the same occurrence of aberration at the second test the 
exercise was allowed to continue. 



Reproducibility of exercise in atrial fibrillation 29 

Table 111. Heart rate resting, standing and during exercise 
A is the first exercise test. B-A is the difference between the second and the first test. W ,  is the first load and Wz is the 
second load. Patients 1-8 have the lowest W,,, and patients 9-15 have the highest W,,,. Definitions in Methods 

Patients 1-8 Patients 9-15 Total 

Mean S.D. Mean S . D .  Mean S.D. 

Resting 
A 
B -A 

A 
B-A 

A 
B-A 

A 
B-A 

W,,, 
A 
B-A 

Standing 

W ,  

W* 

75.0 
2.1 

10.4 
-1.6 

104.8 
4.9 

125. I 
5.8 

129.4 
10.3 

15.3 
4.0 

9.9 
6.7 

19.0 
10.9 

25.4 
12.0 

22.1 
11.6 

77.4 
-7.3 

17.3 
-0.2 

107.6 
-6.8 

134.0 
-5.4 

154.2 
-4.3 

9.8 
6.7 

5.6 
3.9 

23.1 
8.4 

26.1 
9.6 

18.1 
9.6 

76.1 
-2.2 

13.6 
-0.9 

105.8 
0.4 

128.5 
1.5 

137.5 
3.1 

12.6 
4.3 

8.7 
5.4 

19.8 
10.0 

25 .o 
11.0 

22.4 
11.3 

were almost identical, 2 had a difference of about 
1/3 and the remaining 2 had about 50% increase 
from the lowest value in the compared pair. 

W,,, values were identical in 13 of the 15 pa- 
tients. In one patient, on the second occasion, the 
maximal load was chosen as 500, due to a mistake, 
instead of as planned 450 kpmlmin, while at lower 
loads that were the same there was a good agree- 
ment between the two tests. In another patient, 
there were similar heart rates at 200 and 400 
kpmlmin but on the first exercise test the in- 
vestigator finished the test due to aberrant veptricu- 
lar ECG-complexes. The second test was contiwed 
in spite of a similar aberration. 

In 3 patients with a W,,, of 500 kpmlmin and 
more the highest attained heart rates were almost 
identical at the two tests, 150 and 146 in one, 171 
and 171 in another and 127 and 123 in the third, 
respectively. In 2 patients who had a W,,, of 400, 
one had heart rates of 141 and 134 but the other 
patient had 163 and 137. Out of the 8 patients with a 
low W,,,, 2 had shown a good agreement between 
the heart rates at the two tests. In Table I11 the dif- 
ferences between the patients with a low and those 
with a higher W,,, are seen. It is apparent from the 
Table that the S.D. of the differences between the 
two exercise tests is smaller in the group with a 
better W,,,. The only exception is at rest where the 
S.D. was higher in the group with a higher W,,,. 

A ‘steady state’ was achieved on both occasions 

in only 2 patients. In 9 patients a ‘steady state’ 
condition at the highest load was obtained during 
one exercise test but not the other. Six of these 
patients obtained ‘steady state’ conditions on the 
second exercise test. 

DISCUSSION 

In many patients the case history is sufficient for 
a clinically appropriate judgment of the working 
capacity. However, i t  is sometimes difficult to get a 
reliable description from the patient and also i t  is 
often necessary to obtain a more objective and pre- 
cise measurement on the functional capacity. 

This is for example necessary to evaluate results 
after medication, surgery, or in an attempt to find 
an optimal time for surgical intervention etc. The 
need of a reliable method in this respect is obvious. 
Such a method is of course to a critical degree 
dependent upon a good reproducibility of the pro- 
cedure. The ideal method should be well 
standardized, easy to repeat and the results should 
not be influenced by the training effect of repetition. 
The method used in the present study probably 
fulfils these requirements. A training effect of any 
significance is not probable considering the time 
lapse between the two tests and the fact that the test 
was repeated only once. 

A most important factor with regard to the 
purpose of this study, i.e. to evaluate the repro- 



30 H .  Aberg et al. 

ducibility of the exercise test is the stability of the 
patients with regard to their cardiac state. Most 
patients had got their atrial fibrillation in the course 
of a progressive heart disease and it was difficult to 
ascertain that the patients were stable enough for 
the purpose of this investigation. 

I n  spite of a careful selection of patients, the 
results of the exercise tests in the patients with a 
low working capacity indicate either that such 
stable conditions were not obtained or that the ar- 
rhythmia was highly variable in itself. In the group 
of patients with a better heart function there was a 
relatively good reproducibility of the orthostatic 
test reaction as well as the exercise test. 

The ventricular heart rate in atrial fibrillation is 
dependent upon the conditions of the atrioventricu- 
lar junction (7, 1 I ) .  It has been shown that the rate 
of the atrial activity is not changed during exercise 
in patients with atrial fibrillation (1). In spite of the 
irregular atrial activity the atrial rate is so rapid that 
there is always an impulse on hand for propagation 
through the atrioventricular junction. Therefore, 
the ventricular response in atrial fibrillation should 
be dependent upon the function of the atrioventricu- 
lar node. Thus, we must consider influences from 
pharmacologic agents and the autonomous nervous 
system as well as from the heart disease in itself. I n  
a previous paper we have shown that a rather high 
digitalis dosage is favourable for patients with atrial 
fibrillation (3). This has also been studied by Red- 
fors (8). 

In this study, however, there was no difference in 
drug therapy or electrolyte balance between the two 
tests. The alteration in atrioventricular function, 
particularly in those individuals with low working 
capacity, should therefore be caused either by vari- 
ation in autonomic tone or by changes in the under- 
lying heart disease. 

In conclusion, the reproducibility of exercise 
tests was acceptable in patients with a good work- 
ing capacity. On the other hand, in the group with 
more severely impaired heart function, the repro- 
ducibility was poor in spite of digitalization. 

REFERENCES 
1. Aberg, H .  & Furberg, B.: Atrial activity during exer- 

cise in patients with atrial flutter or atrial fibrillation. 
Ups J Med Sci80: 20, 1975. 

2. Aberg, H . ,  Strom, G. & W e r n e r ,  I.: Heart rate during 
exercise in patients with atrial fibrillation. Acta Med 
Scand 191; 3 15, 1972. 

3 .  

4. 

5. 

6. 

7. 

8. 

- The effect of digitalis on the heart rate during 
exercise in patients with atrial fibrillation. Acta Med 
Scand 191:441, 1972. 
Holmgren, A , ,  Jonsson, B . ,  Linderholm, H . ,  
Sjostrand, T. & Strom, G . :  Physical working capacity 
in cases of mitral valvular disease in relation to heart 
volume, total amount of hemoglobin and stroke vol- 
ume. Acta Med Scand 167: 99, 1958. 
Holmgren, A. & Mattsson, K .  H . :  A new ergometer 
with constant load at varying pedalling rate. Scand J 
Clin Lab Invest6: 137, 1954. 
Holmgren, A .  & Strandell, T.: On the use of 
chesthead leads for recording of electrocardiogram 
during exercise. Acta Med Scand 169: 57, 1961. 
Know, J. A .  C.: The heart rate with exercise in pa- 
tients with auricular fibrillation. Br Heart J ll: 119, 
I 949. 
Redfors, A , :  Digoxinbehandling vid formaksflirnmer. 
Relationen mellan dos, plasmakoncentration och ef- 
fekt. Akademisk avhandline., Lund 1971. 

I 

9. Sandberg, L.: Studies on electrocardiographic 
changes during exercise tests. Acta Med Scand 
[Suppl.] 36.5, 1961. 

10. Sjostrand, T.: Changes in the respiratory organs of 
workmen at an ore smelting works. Acta Med Scand 
[Suppl.] 196: 687, 1947. 

1 1 .  Soderstrom, N.: What is the reason for the ventricular 
arrhythmia in cases of auricular fibrillation? Am 
Heart J40:212, 1950. 

12. Varnauskas, E., Cramkr, G . ,  Malmcrona, R., Dahl, 
L.-E., Nystrom, B., Wassen, A. & Werko, L.: Res- 
toration of normal sinus rhythm in patients with 
mitral-valve disease and atrial fibrillation. Nord Med 
62: 1109, 1959. 

13. Wahlund, H.: Determination of the physical working 
capacity. Acta Med Scand [Suppl.]215, 1948. 

Received September 2, 1976 

Address for reprints: 

H a n s  Aberg, M.D. 
Department of Internal Medicine 
University Hospital 
S-750 14 Uppsala 14 
Sweden 

Upsula .I M e d  Sci 82